bath and the inset beakers). We recommend the following ultrasonic products: 
STAMMOPUR DR 8, STAMMOPUR R, TICKOMED 1, TICKOPUR TR 3, TICKOPUR R 30, 
TICKOPUR TR 7

If none of these products are available, a neutral or mildly alkaline product that does not 
destroy aluminium may be used. The product must be approved by the manufacturer for 
use in ultrasonic baths. 

Conducting the foil test: 

Commercially available aluminium foil, scissors, a waterproof pen, a sample holder and 
associated inset beakers are required to perform the foil test. In the case of several foil 
tests carried out, a comparison with previously produced foils follows after implementation, 
or foil generated during initial commissioning. 

IMPORTANT

The foil test must be conducted in inset beakers since aluminium foil remains may clog 
the cooling circuit. Therefore, it must always be ensured that no aluminium foil gets into 
the ultrasonic bath, even during the foil test. 

1.  Fill 2/3 of the ultrasonic bath with water and 1 % of a suitable preparation 

(= 99 parts water + 1 part concentrate), see chapter 4.1.

2.   Degas liquid

3.   Also 

fi

 ll inset beakers with the liquid described under 1.

4.  Fill the inset beakers with the same liquid, hang them inside the ultrasonic bath and 

also degas them for approx. 5 minutes

5.   Switch o

  the ultrasound

6.   Cut aluminium foil (household foil 10 

μ

m to 25 

μ

m thick) evenly so that the parts 

fi

 t into 

the inset beakers

7.   Submerge one foil cut per inset beaker

8.   Switch on ultrasound, sonicate for 1 minute at least until there is visible perforation 

or hole formation

9.   If necessary, increase sonication time up to 3 minutes for more stable foils 

(thicker or coated) 

10. Switch o

  the ultrasound

11.  Take a note of the position of the individual foils in the containers or document 

with photos

12. Switch o

  the ultrasound, take the foil out and let it dry

13. Label the foils accordingly with the position of the inset beaker, 

duration and date, and then archive

14. Rinse inset beakers thoroughly to 

remove detached foil particles

15. Check ultrasonic bath for foil residue, 

if necessary renew and 

fl

 ush all liquid, 

see chapter 4.4.

  Perforation of the foil must be visible 

(see illustration), otherwise contact 
the manufacturer

 Inset beaker

Aluminium foil, 

must be immersed

Fill level in 

the inset 

beaker

Sample slide from 
an inset beaker

Содержание SONOCOOL 255.2

Страница 1: ...Operating Instructions 51806 001 en 2021 12 C High power ultrasonic bath SONOCOOL 255 2...

Страница 2: ...he original version will take precedence BANDELIN accepts no responsibility or liability for damage caused by improper handling or usage contrary to the intended purpose The documentation has been pre...

Страница 3: ...s both the warranty claim and the CE conformity will become void If service is required please contact the authorised dealer or the manufacturer Symbols used Symbol Meaning Explanation Danger Denotes...

Страница 4: ...elements 14 3 1 1 Turning the ultrasonic bath on o 14 3 1 2 Explanation of the display fields 15 3 2 Signals on the ultrasonic bath 15 3 3 Operating modes 16 3 4 Setting the operating parameters 17 3...

Страница 5: ...4 Check the cooling 30 6 1 5 Checking the ultrasound e ect 30 6 2 Maintenance by manufacturer or qualified personnel 32 6 2 1 Visual inspection 33 6 2 2 Functional check 34 6 2 3 Pre cleaning 34 6 2 4...

Страница 6: ...e plate Product features Stainless steel oscillating tank 1 with oscillating systems ultrasonic frequency 35 kHz Filling level mark for safe filling 2 electronic fill level sensor 2a Operating and dis...

Страница 7: ...51806 001 en 2021 12 7 43 1 electronic GmbH Co KG Heinrichstra e 3 4 12207 Berlin Deutschland info bandelin com 11 9 10 2 2a 7 6 1...

Страница 8: ...laboratories or research facilities 1 2 Mode of operation With the bath it is possible to use the catalytic e ect of ultrasound for processes during which simultaneous cooling is required Heat sensit...

Страница 9: ...404 2 mm welded Working fill capacity 5 litres contact liquid 4 litres in the tank 1 litre in the cooling unit Adjustable bath temperature 4 40 C in 20 C environment Refrigerant R290 Refrigerant quant...

Страница 10: ...instructed in its operation by reference to these instructions We will not o er any guarantee for damages to the ultrasonic bath or oscillating tank or to the objects to be treated as a result of use...

Страница 11: ...s do not refresh them by adding more liquids Follow the manufacturer s dosage instructions Cooling Information on the cooling unit in the SONOCOOL ultrasonic bath Refrigerant R290 quantity 90 g extrem...

Страница 12: ...ocation so that it can adjust to the ambient conditions 2 1 Scope of delivery Set for pathology order no 3500031 comprising 1 ultrasonic bath with integrated cooling unit order No 350003 including 1 s...

Страница 13: ...horoughly rinse the ultrasonic bath oscillating tank with water before its first use Note All outer surfaces and also the inner walls of the oscillating tank are covered with an oily preservative in o...

Страница 14: ...turning on the LCD display must light up Initialisation occurs automatically BANDELIN electronic S O N O C O O L initialising Next the ultrasonic bath will be on stand by mode and will be ready for op...

Страница 15: ...in the oscillating tank in C Target runtime Specified value for the duration of the ultrasound operation in hh mm ss format Note As soon as the ultrasonic bath is turned on the cooling will run automa...

Страница 16: ...d parameter and the current value are shown on the top portion of the display All valid target values are shown on the bottom rows Stand by mode SET key Active mode SET key Pause mode SET key Active m...

Страница 17: ...e in editing mode Setting limits 4 40 C Example taken from stand by mode Editing mode temperature 12 00 00 17 0 C 50 Temperature 17 2 C 17 12 00 00 17 0 C 50 12 00 00 17 2 C 12 00 00 SET Notes If the...

Страница 18: ...limits 00 00 10 99 59 59 hh mm ss Example taken from stand by mode Editing mode temperature Editing mode temperature 12 00 00 17 0 C 50 Temperature 17 2 C 17 12 00 00 17 0 C 50 Temperature 17 2 C 17 1...

Страница 19: ...time by pressing the START STOP key once again in pause mode By pressing the SET key for 2 sec the procedure can be interrupted from pause mode and the display values can be reset return to stand by m...

Страница 20: ...3 For additional information see chapter 4 5 Notes Filling Do not fill the oscillating tank with hot water 30 C The filling temperature should be between 15 and 25 C Water of at least drinking qualit...

Страница 21: ...accessories from the oscillating tank Turn on the ultrasonic bath The internal pump starts pumping the cooling runs automatically according to the set target temperature Note If cooling is not desired...

Страница 22: ...e inset beakers are immersed a minimum of 2 cm in the contact liquid Possible air bubbles are completely removed from below the vessels When using sample containers e g sample cassettes for decalcific...

Страница 23: ...on the rear side Remove the protective cover from the emptying nozzle and slide the supplied drain hose onto the emptying nozzle Place the other end of the drain hose in a suitable recipient or in th...

Страница 24: ...s not form in the entire solution but rather only on the oscillating tank bottom or in water inhomogeneities in general To reduce the surface tension add a surfactant preparation that is biodegradable...

Страница 25: ...aqueous agents made by DR H STAMM GmbH are prepared pursuant to the regulations of the German Washing and Cleansing Agents Act are biodegradable and as working solutions may be disposed of in the wast...

Страница 26: ...t the ultrasonic bath see chapter 4 4 Repeat the described rinsing process two to three times After the last emptying wipe dry the oscillating tank using a soft cloth Oscillating tank The oscillating...

Страница 27: ...of the bath and around the outflow which can result in a cross infection Consequently the baths and surfaces of the device must be disinfected and cleaned regularly The disinfection should be perform...

Страница 28: ...flushed at regular intervals to prevent possible deposits resulting from the circulating contact liquid To do so check whether the flow is at approx 3 5 l min TICKOPUR TR 13 is recommended in 5 dosage...

Страница 29: ...c bath with the appropriate contact liquid if recommissioning is planned If impurities still come out of the tube system during rinsing it is recommended that the tube system is flushed again If the f...

Страница 30: ...sibility of the user The foil test is a simple procedure for demonstrating the intensity and distribution of cavitation in an ultrasonic bath For this purpose an aluminium foil stretched on a foil tes...

Страница 31: ...le preparation 99 parts water 1 part concentrate see chapter 4 1 2 Degas liquid 3 Also fill inset beakers with the liquid described under 1 4 Fill the inset beakers with the same liquid hang them insi...

Страница 32: ...s the safety checks Maintenance kit Scope of delivery Maintenance kit order no 350003100 1 T hose connector 3 L hose connector 1 cooling water arch 1 insulating hose 12 9 mm 2 m long to be cut to leng...

Страница 33: ...fications Until the ultrasonic bath has been disinfected protective clothing e g disposable gloves Focus of disinfection Oscillating tank Pump Refrigeration unit Outlet Hoses inside When replacing com...

Страница 34: ...onic bath with TR 13 10 4 5 l water 50 ml TR 13 Switch on the ultrasonic bath Operate the ultrasonic bath without cooling for 1 hour set target temperature to 40 C cooling cannot be switched o otherwi...

Страница 35: ...te of the pump is less than 3 5 l min the pump must be cleaned or replaced order no 3500160 Rinse the ultrasonic bath with TICKOMED 1 3 4 85 l water 150 ml TICKOMED 1 Switch on the ultrasonic bath Ope...

Страница 36: ...ultrasound cannot be started and or the ultrasonic bath turns o for safety reasons Wait until the cooling has su ciently cooled the contact liquid tolerance range target temperature 1 C 012 Error 012...

Страница 37: ...er 6 1 2 Circulation pump defective Check whether there is bubbling in the tank If not please contact the manufacturer Unsatisfactory results Degassing not conducted Always conduct degassing during th...

Страница 38: ...repair The ultrasonic bath does not need to be opened in order to replace the fuses The fuses 2 x microfuses 5 20 mm T4A are housed in a plug in module in the mains socket which is located on the rear...

Страница 39: ...the manufacturer must be contacted BANDELIN electronic GmbH Co KG Heinrichstrasse 3 4 12207 Berlin Germany Repair service E mail Tel 49 0 30 768 80 13 info bandelin com Fax 49 0 30 76 88 02 00 13 Retu...

Страница 40: ...nic application easier while protecting the oscillating tank and the objects to be treated BANDELIN o ers a wide range of accessories Additional information may be obtained from our supplier our sales...

Страница 41: ...515 Sample holder PH 255 1 3519 Sample holder PH 255 2 3518 Sample holder PH 255 11 3512 Inset beaker SD 01 2 Pkg 10 units 3517 Implant box IB 18 Pkg 5 units 3283 Glass beaker SD 06 Pkg 1 unit 330 Ins...

Страница 42: ...of the reach of children and also of persons who have not been instructed in their use by reference to the product information Do not ingest or inhale the agents and do not allow them to come into con...

Страница 43: ...yword index not applicable BANDELIN electronic GmbH Co KG Heinrichstra e 3 4 12207 Berlin Deutschland www bandelin com info bandelin com 49 30 768 80 0 49 30 773 46 99 Zertifiziert nach ISO 9001 ISO 1...

Отзывы: