51806-001 en/2021-12
11 / 43
1
electronic GmbH & Co. KG • Heinrichstraße 3-4 • 12207 Berlin • Deutschland • [email protected]
Operation
• Observe ambient and set-up conditions, see Section 1.4.
• Only plug in the ultrasonic bath to an outlet with a grounded socket.
• Do not operate the ultrasonic bath without liquids.
• Do
not
fi
ll the oscillating tank with hot water (> 30 °C). The
fi
lling temperature should
be between 15 and 25 °C.
• Do not stand or lay any objects directly on the tank bottom. Use accessories,
see chapter 7.
• Do not immerse any parts of the body (e.g. hands, feet) or living organisms (animals
or plants) in the tank; in particular, do not make contact with the bath liquid during
ultrasound operation. Danger: Ultrasound has a cell-destroying e
ff
ect.
• In the event of continuous activity within a 2 m radius, adequate hearing protection must
be used. Danger: Hearing disturbances during operation are possible if not wearing
hearing protection – the typical ultrasound cavitation noise can be very uncomfortable.
• Fully replace used liquids, do not refresh them by adding more liquids. Follow
the manufacturer's dosage instructions.
Cooling
• Information
on
the cooling unit in the SONOCOOL ultrasonic bath:
-
Refrigerant: R290; quantity: 90 g, extremely
fl
ammable
• The
cooling circuit must not be damaged!
The cooling unit contains
fl
ammable coolant!
If the cooling circuit is damaged, immediately disconnect the power plug and ventilate
the installation room well. Send the ultrasonic bath for repair to the manufacturer.
• Never cover or close the incoming or outgoing air openings on the ultrasonic bath.
• In order to protect the refrigeration unit, the compressor will only start after a resting
period of two minutes.
Damage
• If damage to the ultrasonic bath is detected, do not connect the ultrasonic bath
to the mains.
• In the event of defects, disconnect the power plug immediately.
•
Repairs must only be conducted by authorised skilled personnel or by the manufacturer.
• Defective parts may only be replaced with original parts.
Service
• Maintenance may only be carried out by quali
fi
ed personnel.
• Before opening, disconnect the device from the mains.
• While maintenance work is being carried out on the pump,
there is a risk of crushing
fi
ngers due to high magnetic forces.