– 5 –

1.  ImPortAnt InformAtIon 

AnD IntenDeD USe

Please read the entire instruction manual carefully prior 
to initial operation to protect yourself and your patients 
from operating errors! this instruction manual contains 
instructions on checks which must be performed by the 
practitioner before initial operation and periodically 
thereafter. the checks are needed to ensure the unit and 
its safety features are fully operational. Please keep this 
instruction manual so that it can be consulted whenever 
necessary.

Intended use

Specific percentages of medical nitrous oxide are added 
to medical oxygen using the mixer. the patient remains 
conscious, but is more relaxed and is no longer afraid of 
the treatment. Various journals and studies have indi-
cated that nitrous oxide sedation up to a maximum of 
70% nitrous oxide constitutes minimal sedation aimed 
at achieving an anxiolytic effect (alleviating anxiety) in 
patients. It is not an anaesthetic. the patient is awake 
and breathing independently and can respond to exter-
nal stimuli. Some patients are not sufficiently sedated 
with a level of 50% nitrous oxide. Accordingly, besides 
the Baldus® touch 50, which can administer maximum 
50% nitrous oxide, there are also the Baldus® touch 60 
and 70 units, which are limited to maximum 60% and 
70% nitrous oxide. trained practitioners can therefore 
decide for themselves the maximum nitrous oxide dose 
they would like to administer. the nitrous oxide concen-
tration is tailored to the individual patient. the Baldus® 
touch 50, 60 and 70 units are designed for installation 
in the Baldus® All-in-one-Cart, for example, or in plas-
terboard. the Baldus® Bag t is also compatible with ni-
trous oxide mixers from other manufacturers designed 
for installation. Breathing bags and mask systems are 
attached to the Bag t.

Important information

the oxygen-nitrous oxide mixer must not be combined 
with anaesthetic equipment or used as anaesthetic 
equipment. Adjusting the settings or system and open-
ing the mixer or Bag t are strictly forbidden, as this could 
cause your patient serious harm. Any alterations to the 
mixer or Bag t will invalidate all guarantees and warran-
ties. 
only accessories approved by Baldus® for the system 
may be used. In case of faults or problems discontinue 
the treatment with nitrous oxide immediately and con-
sult your instruction manual and/or contact your ser-

vice partner or the manufacturer Baldus medizintechnik 
GmbH (0049 261/ 96 38 926-0). Do not attempt to per-
form any repairs yourself or adjust the mixer.  never use 
the unit without first taking an advanced training course 
on how to handle and use these units. ensure to careful-
ly observe the content of the advanced training course 
from which you obtained your certification and instruc-
tions from your service partner. the unit is only intended 
for use in the medical field and may only be used by a 
physician to sedate patients with nitrous oxide and ox-
ygen where necessary. the oxygen-nitrous oxide mixer 
must not be used for any other purpose.
Please take the medical history correctly, as conveyed 
during the advanced training course. At the start of the 
treatment, apply the pulse oximeter and increase the 
concentration of n

2

o slowly, according to the individual 

patient, until the patient is sedated. never leave a sedat-
ed patient unattended! the breathing bag can be used 
to monitor the sedation and should always be filled dur-
ing the sedation (can be ordered as an accessory, refer to 
Section 4. Accessories and 7. Handling).

CautIon!  Any sedation must always be monitored 
using a pulse oximeter with individual alarm limits 
(refer to Section 4. Accessories)! Always keep spare 
batteries ready for the pulse oximeter! 

Pulse oximeter

 (can be ordered as an accessory, refer 

to Section 4. Accessories): the pulse and oxygen satura-
tion level are vital parameters which indicate whether 
the patient is doing well and the sedation is having an 
effect. Sedation usually increases the oxygen saturation 
level, lowers the pulse and relaxes the patient. If the ox-
ygen saturation level drops below an individually ad-
justable value, an alarm sounds on the pulse oximeter. 
If a patient is being sedated with 20% nitrous oxide, for 
example, 80% oxygen is being delivered. In other words, 
the patient has a healthy oversupply of oxygen. If the 
oxygen saturation level still falls below 85%, for exam-
ple, the nitrous oxide sedation must be stopped imme-
diately in case the nitrous oxide and oxygen have been 
mixed up and 80% nitrous oxide is being delivered, for 
instance. ensure the oxygen and nitrous oxide have not 
been mixed up (refer to Section 6.1 General safety infor-
mation: gas type check).
Delivering 100% nitrous oxide would harm your patient. 
However, there is very little probability of this happening 
as the manufacturer performs a gas type check. If the pa-

Содержание Touch 50

Страница 1: ...0483 Made in Germany Instructions for use 02 N2 0 Mixer for Analgesia Baldus Touch 50 60 70 Baldus Bag T...

Страница 2: ...sedation and medical gases Welcome to Baldus Medizintechnik GmbH Baldus Medizintechnik GmbH Auf dem Schafstall 5 56182 Urbar Germany 49 0 261 96 38 926 0 49 0 261 96 38 926 21 office baldus medizintec...

Страница 3: ...g T 11 3 5 Back and side views 11 3 6 Technical specifications 13 4 Scope of delivery and accessories 14 5 Assembly instructions 15 6 Safety and regular controls 18 6 1 General safety instructions 18...

Страница 4: ...ed 01 06 2017 0483 Inhalt nicht steril Contents not sterile Physicians using nitrous oxide have given our service a 1 1 4 rating You can trust the specialist in medical gases and inhalation sedation s...

Страница 5: ...and con sult your instruction manual and or contact your ser vice partner or the manufacturer Baldus Medizintechnik GmbH 0049 261 96 38 926 0 Do not attempt to per form any repairs yourself or adjust...

Страница 6: ...are facilities Take the table under Section 6 on reduc ing ambient air pollution through nitrous oxide into ac count A suction unit is compulsory and should have a suction power of at least 45 l min I...

Страница 7: ...effects The general side effects of nitrous oxide are very minor e g vitamin B12 deficiency In case of continuous admin istration over two days Cardiovascular side effects No relevant side effects fr...

Страница 8: ...te When creating a medical history please follow the content of the advanced training course Oxygen nitrous oxide mixer Controls and regulates the flow gas flow rate and per centage nitrous gas and ox...

Страница 9: ...switched at any time 3 2 Main menu 1 Start sedation with patient ID A menu with letters and numbers appears for adding a patient by name or patient number After clicking Enter the sedation starts imm...

Страница 10: ...last 20 error messages of the unit can be viewed and saved The memory can also be deleted All errors are displayed in the maintenance menu 4 7 Recovery time The recovery time can be set here individu...

Страница 11: ...eathe in pure oxygen using the full face mask The demand valve of the mask opens on inhalation If a patient is unconscious and no longer breathing independently he can be actively ventilated by pressi...

Страница 12: ...ormat FAT32 Possible sizes of up to 64 GB Fig 7 Baldus Touch view of the left side Fig 8 The full Baldus Touch system in the Baldus All in One Cart BACK FRONT All in One Cart N2 O pressure hose O2 pre...

Страница 13: ...ation 0 Maximum N2 O concentration 60 Product variant 3 Baldus Touch 50 for installation Minimum O2 concentration 50 Maximum O2 concentration 100 Minimum N2 O concentration 0 Maximum N2 O concentratio...

Страница 14: ...cer 16038 1 O2 Pressure Hose 16038 2 N2 O Pressure Hose 15015 3L or 15015 2 3L 3 L Breathing Bag for Adults or 2 3 L Breathing Bag for Children 15016 Baldus Holder for Cylinders e g cylinder trolley 1...

Страница 15: ...keep a spare oxygen cylinder on hand For a centralised gas supply the gas cylinders and equip ment are located in a separate room and the gas flows to the treatment rooms through medical copper pipes...

Страница 16: ...e used for this As the connectors are different sizes there is no possibility of confusion Assembling the Baldus Touch mixer and Baldus Bag T The Baldus Touch oxygen nitrous oxide mixer is in stalled...

Страница 17: ...here 2 2 Now go from the point for the nozzle out 15 mm to the right and 15 mm upwards and draw the second point for the screws there 2 3 Now go from the point for the nozzle out 15 mm to the right an...

Страница 18: ...ns and the Or dinance on Hazardous Substances Gefahrstoffverord nung The gas types must be checked after installing the mixer and accessories to rule out any mix up in the gas types To check the gas t...

Страница 19: ...tage administer 100 oxygen to your patient to discontinue the nitrous oxide sedation and start the recovery phase If you do not have any ox ygen supply due to the power outage remove the mask from the...

Страница 20: ...ow or there is a measure ment section failure The visual Check O2 signal appears in the top right of the sedation display and the O2 bar on the bottom right is shaded red A Check O2 gas supply pop up...

Страница 21: ...an cause malfunctions Have annual safety checks and maintenance per formed including filter checks Service staff employees or partners of Baldus Mediz intechnik GmbH can use a service USB stick to acc...

Страница 22: ...Leakscanbeeasilycheckedforusingaspecialleakdetec tion spray which is sprayed on all screw joints and con nections on the mixer and pressure reducers The leak detection spray consists of soap and wate...

Страница 23: ...rs 5 Now close the oxygen cylinder 6 The Check O2 information signal appears after a short period of time 7 The system is vented Both pressure reducers now show 0 bar 8 You can now switch off the unit...

Страница 24: ...n der is changed and at least monthly Close the gas cylinders completely there should be no appreciable drop in pressure for an hour Always open the high pressure valves of the gas cylinder s slowly O...

Страница 25: ...the left concentration controller to add nitrous oxide to the mix as learned in the advanced training course Titrate the nitrous oxide individually for each patient To end the nitrous oxide sedation p...

Страница 26: ...ople used to be administered a 100 dose of nitrous oxide at carnivals for fun after which they would stagger around the place comically This would make onlookers laugh j Can I sedate a patient with a...

Страница 27: ...lable allen anwendbaren Anforderungen der Richtlinie 93 42 EWG Anhang II entspricht meets all applicable requirements of the Directive 93 42 EEC Annex II Name Adresse und Kennnummer mdc medical device...

Страница 28: ...Baldus Medizintechnik GmbH Auf dem Schafstall 5 56182 Urbar Germany 49 0 261 96 38 926 0 49 0 261 96 38 926 21 office baldus medizintechnik de www baldus medical com...

Отзывы: