3.6 technical data

PhySICaL FEatuRES 
Mixer:

dimensions:

 

135 x 300 x 120 mm (H x W x D)

Weight:

 

2.5 kg 

nitrous oxide/n

2

o inlet:

  DISS connector

oxygen-o

2

 inlet:

 

DISS connector

Bag t:

diameter of the bag and  
scavenger system connectors:

 

ID 15 mm x OD 22 mm 30 x 40 x 70 mm (L x W x D)

Back of the Bag T:

 hose nozzle, 40 x 8 mm (L x W)

Gas administration 

Maximum flow: 

18 l/min.

demand mask connector: 

100 to 250 l/min.

PRoDuCt VaRIant 1:  
Baldus® touch 70 (for installation)

minimum o

2

 concentration: 30%

maximum o

2

 concentration: 100%

minimum n

2

o concentration: 0%

maximum n

2

o concentration: 70%

PRoDuCt VaRIant 2:  
Baldus® touch 60 (for installation)

minimum o

2

 concentration: 40%

maximum o

2

 concentration: 100%

minimum n

2

o concentration: 0%

maximum n

2

o concentration: 60%

PRoDuCt VaRIant 3:  
Baldus® touch 50 (for installation)

minimum o

2

 concentration: 50%

maximum o

2

 concentration: 100%

minimum n

2

o concentration: 0%

maximum n

2

o concentration: 50%

PRoDuCt VaRIant 4: 
Baldus® Bag t

Compatible with Baldus® touch 50/60/70, matrx mDm-D 
Cabinet mount Digital flowmeter, Porter Digital mDm, 
Porter mxr-1 Cabinet-mounted flowmeter, Porter 
mxr-D Digital flowmeter and Accutron Digital Ultra™ 
flushmount nitrous oxide unit    

Fig. 10: Product variant 4: Baldus® Bag t

Fig. 9: Product variant 1-3: Baldus® touch 
(for installation) 

Electrics

Input voltage: 110/230V AC
Input frequency: 50 – 60 Hz
fuses: 2 x 1A, inert
max. ambient temperature: 50°C
Power pack:
Protection class: I
Ampere: 1.0-0.55A

Inlet pressure 

Inlet pressure: min. 3.5, max. 5 bar
the outlet pressure is less than 0.3 bar. 

note: All specifications are subject to tolerances 
during manufacture.

non-continuous operation

Содержание Touch 50

Страница 1: ...0483 Made in Germany Instructions for use 02 N2 0 Mixer for Analgesia Baldus Touch 50 60 70 Baldus Bag T...

Страница 2: ...sedation and medical gases Welcome to Baldus Medizintechnik GmbH Baldus Medizintechnik GmbH Auf dem Schafstall 5 56182 Urbar Germany 49 0 261 96 38 926 0 49 0 261 96 38 926 21 office baldus medizintec...

Страница 3: ...g T 11 3 5 Back and side views 11 3 6 Technical specifications 13 4 Scope of delivery and accessories 14 5 Assembly instructions 15 6 Safety and regular controls 18 6 1 General safety instructions 18...

Страница 4: ...ed 01 06 2017 0483 Inhalt nicht steril Contents not sterile Physicians using nitrous oxide have given our service a 1 1 4 rating You can trust the specialist in medical gases and inhalation sedation s...

Страница 5: ...and con sult your instruction manual and or contact your ser vice partner or the manufacturer Baldus Medizintechnik GmbH 0049 261 96 38 926 0 Do not attempt to per form any repairs yourself or adjust...

Страница 6: ...are facilities Take the table under Section 6 on reduc ing ambient air pollution through nitrous oxide into ac count A suction unit is compulsory and should have a suction power of at least 45 l min I...

Страница 7: ...effects The general side effects of nitrous oxide are very minor e g vitamin B12 deficiency In case of continuous admin istration over two days Cardiovascular side effects No relevant side effects fr...

Страница 8: ...te When creating a medical history please follow the content of the advanced training course Oxygen nitrous oxide mixer Controls and regulates the flow gas flow rate and per centage nitrous gas and ox...

Страница 9: ...switched at any time 3 2 Main menu 1 Start sedation with patient ID A menu with letters and numbers appears for adding a patient by name or patient number After clicking Enter the sedation starts imm...

Страница 10: ...last 20 error messages of the unit can be viewed and saved The memory can also be deleted All errors are displayed in the maintenance menu 4 7 Recovery time The recovery time can be set here individu...

Страница 11: ...eathe in pure oxygen using the full face mask The demand valve of the mask opens on inhalation If a patient is unconscious and no longer breathing independently he can be actively ventilated by pressi...

Страница 12: ...ormat FAT32 Possible sizes of up to 64 GB Fig 7 Baldus Touch view of the left side Fig 8 The full Baldus Touch system in the Baldus All in One Cart BACK FRONT All in One Cart N2 O pressure hose O2 pre...

Страница 13: ...ation 0 Maximum N2 O concentration 60 Product variant 3 Baldus Touch 50 for installation Minimum O2 concentration 50 Maximum O2 concentration 100 Minimum N2 O concentration 0 Maximum N2 O concentratio...

Страница 14: ...cer 16038 1 O2 Pressure Hose 16038 2 N2 O Pressure Hose 15015 3L or 15015 2 3L 3 L Breathing Bag for Adults or 2 3 L Breathing Bag for Children 15016 Baldus Holder for Cylinders e g cylinder trolley 1...

Страница 15: ...keep a spare oxygen cylinder on hand For a centralised gas supply the gas cylinders and equip ment are located in a separate room and the gas flows to the treatment rooms through medical copper pipes...

Страница 16: ...e used for this As the connectors are different sizes there is no possibility of confusion Assembling the Baldus Touch mixer and Baldus Bag T The Baldus Touch oxygen nitrous oxide mixer is in stalled...

Страница 17: ...here 2 2 Now go from the point for the nozzle out 15 mm to the right and 15 mm upwards and draw the second point for the screws there 2 3 Now go from the point for the nozzle out 15 mm to the right an...

Страница 18: ...ns and the Or dinance on Hazardous Substances Gefahrstoffverord nung The gas types must be checked after installing the mixer and accessories to rule out any mix up in the gas types To check the gas t...

Страница 19: ...tage administer 100 oxygen to your patient to discontinue the nitrous oxide sedation and start the recovery phase If you do not have any ox ygen supply due to the power outage remove the mask from the...

Страница 20: ...ow or there is a measure ment section failure The visual Check O2 signal appears in the top right of the sedation display and the O2 bar on the bottom right is shaded red A Check O2 gas supply pop up...

Страница 21: ...an cause malfunctions Have annual safety checks and maintenance per formed including filter checks Service staff employees or partners of Baldus Mediz intechnik GmbH can use a service USB stick to acc...

Страница 22: ...Leakscanbeeasilycheckedforusingaspecialleakdetec tion spray which is sprayed on all screw joints and con nections on the mixer and pressure reducers The leak detection spray consists of soap and wate...

Страница 23: ...rs 5 Now close the oxygen cylinder 6 The Check O2 information signal appears after a short period of time 7 The system is vented Both pressure reducers now show 0 bar 8 You can now switch off the unit...

Страница 24: ...n der is changed and at least monthly Close the gas cylinders completely there should be no appreciable drop in pressure for an hour Always open the high pressure valves of the gas cylinder s slowly O...

Страница 25: ...the left concentration controller to add nitrous oxide to the mix as learned in the advanced training course Titrate the nitrous oxide individually for each patient To end the nitrous oxide sedation p...

Страница 26: ...ople used to be administered a 100 dose of nitrous oxide at carnivals for fun after which they would stagger around the place comically This would make onlookers laugh j Can I sedate a patient with a...

Страница 27: ...lable allen anwendbaren Anforderungen der Richtlinie 93 42 EWG Anhang II entspricht meets all applicable requirements of the Directive 93 42 EEC Annex II Name Adresse und Kennnummer mdc medical device...

Страница 28: ...Baldus Medizintechnik GmbH Auf dem Schafstall 5 56182 Urbar Germany 49 0 261 96 38 926 0 49 0 261 96 38 926 21 office baldus medizintechnik de www baldus medical com...

Отзывы: