incorporated into the design to improve ergonomics so the operator can use the cabinet properly without 
sacrificing comfort or safety. 

 

 

Motor/ blower capacity 

A motor/blower's capacity is measured by its ability to provide a nearly constant volume of air as 
resistance increases because of filter loading. Verification by a simulated filter-loading test has 
established that the SterilGARD

III cabinet is capable of automatically handling a 50% increase in 

pressure drop across the filter without reducing total air delivery more than 10%. With the use of the 
manual speed control, a 200% increase in the pressure drop across the filter can be handled. This results 
in extended filter life, which relates to significant cost savings over the life of the cabinet. 

 

One-piece interior wall construction 

The interior side and rear walls of the work area are constructed from a single piece of 16-gauge 
stainless steel. It has 7/16” [11.1mm] radius (rounded) corners to help prevent buildup of contaminants 
and aid in cleanup. 

 

Protective screen 

Located under the work surface at the bottom rear and sides of the return-air plenums, a protective 
screen, integral to the interior walls, is provided to prevent wipes and other paper materials from being 
drawn into the blower system. This feature eliminates costly decontamination processes and downtime.  
The screen affords easy access to trapped materials and should be kept clean at all times. 

 

All-metal plenums 

The plenums are constructed entirely of either carbon steel or stainless steel in order to provide strength, 
durability, air-tightness and resistance to deterioration. 

 

Recessed stainless steel work surface 

The work surface is constructed of corrosion-resistant 16-gauge type 304 stainless steel, with a  satin 
finish that diminishes light reflection.  It is recessed to retain spills, and the radiused corners make for 
easy cleaning and disinfection. 

 

Cabinet exterior panels 

External panel construction is of 16-gauge cold-rolled steel, or 16-guage stainless steel protected by a 
smooth white baked enamel finish. 

 

Tested HEPA filters 

Both the downflow and exhaust filters in the SterilGARD

III Advance

 are scan-tested HEPA filters. 

They are 99.99% effective on particles of 0.3 micron size. Each filter is scan-tested after installation to 
assure that there are no leaks >.01% of upstream concentration. 

 

Viewscreen  

The cabinet's ¼” safety plate glass sliding viewscreen may be opened to 18-¼ ” for placement of large 
items in the work area, and may be fully closed for system shutdown or UV light operation. 

 

Содержание SterilGARD SG303

Страница 1: ...ODEL SG303 SG303 INT SG303 JPN SG303 AUS Revision Date Description F 3 25 15 Updated branding and split manuals This manual includes information for installation operation maintenance and spare parts...

Страница 2: ...ue culturing of possibly infectious samples but also for I V drug preparations and other pharmaceuticals that could have adverse health effects on operators and other techniques requiring a contaminat...

Страница 3: ...djustable feet optional 6 Telescoping stand with castors optional 6 Solid state electronic ballast 7 Programmable Logic Controller 7 UniPressure Plenum 7 Ground Fault Circuit Interrupter 7 Ultraviolet...

Страница 4: ...GRITY TEST 23 ELECTRICAL SAFETY TESTS 24 MAINTENANCE NOTES 24 Cleaning the Work Area 24 Ultraviolet Germicidal Lamp optional 24 Check the Mag Gauge Or Optional Air Flow Monitor AFM Regularly 24 Replac...

Страница 5: ...and the work surface Turbulence caused by friction will also be found along a cabinet s side walls In cabinets without high velocity return air slots this turbulence may also allow contaminants to esc...

Страница 6: ...rear wall the negative pressure will create suction and pull air in not allowing it to escape into the operator s area If there is a leak from the positive pressure plenum the surrounding negative pre...

Страница 7: ...ls is provided to prevent wipes and other paper materials from being drawn into the blower system This feature eliminates costly decontamination processes and downtime The screen affords easy access t...

Страница 8: ...take imbalance due to pressure resistance drop differences of supply and exhaust filters if they are changed at a later date Easy filter access For convenience and ease of service the downflow and exh...

Страница 9: ...lter life and reducing noise 2 The two piece plenum serves as the filter sealing mechanism providing a continuous and evenly distributed clamping force around the complete perimeter of the filters Thi...

Страница 10: ...ularly Specifications Weight The weight of the SterilGARD III Model SG303 cabinet is 397 pounds with a shipping weight of 534 pounds With the optional stand the cabinet weight is 457 pounds with a shi...

Страница 11: ...ith a forklift or other available equipment If any concealed damage is found it should be reported to the delivering carrier A claim for restitution should be filed within 15 days Due to the risk of m...

Страница 12: ...the laboratory instead of venting to the outside it is important that there is adequate space between the top of the cabinet and the ceiling A solid ceiling located too close to the exhaust filter wi...

Страница 13: ...only particulates and not gases thus causing the recirculation of these within the cabinet Ensure that the exhaust filter guard is properly installed over the filter opening The filter guard will prov...

Страница 14: ...net However if the risk is professionally evaluated and a decision is made to install a flammable gas petcock then an emergency shut off valve should be situated in an accessible location outside the...

Страница 15: ...NFORMATION For a complete listing of articles papers and reports related to containment clean air products and safety contact The Baker Company for our complete Bibliography or visit our website at ww...

Страница 16: ...erefore it is recommended that the blower is left on at all times Operation and Controls The operator controls with indicators are arranged on the front electrical panel of the cabinet A number of swi...

Страница 17: ...creen must be placed at the allowable opening of 8 203 2mm The sash alarm will be activated whenever the viewscreen is at a potentially unsafe position To mute the audible alarm press the alarm mute b...

Страница 18: ...y the front of the cabinet while it is in use 4 Avoid using floor type pipette discard canisters It is important that your used pipettes be discarded into a tray or other suitable container inside the...

Страница 19: ...good reference for this procedure is NSF ANSI Standard 49 2002 ANNEX G Recommended Microbiological decontamination Procedure NSF International 789 North Dixboro Road P O Box 130140 Ann Arbor Michigan...

Страница 20: ...et is vented to the outside then the exhaust transition should have a means to recirculate the exhaust airflow back to the cabinet blower This will promote the even distribution of formaldehyde gas th...

Страница 21: ...sinfect the entire inside surface of the cabinet For additional information on cleaning and disinfecting stainless steel please refer to Decontamination Sterilization Disinfection and Antisepsis Vesle...

Страница 22: ...ions you can expect at least five years of use However misuse or a heavy dust load within the cabinet will shorten any filter s lifetime Bunsen burners and misuse of chemicals will also shorten the us...

Страница 23: ...s Your test procedures should be identical to ours so that achieving similar test results and comparison of data will be possible Please correspond directly with us to request detailed procedures for...

Страница 24: ...ve to increase the blower speed in order to maintain the original volume of recirculating air There is a speed control located in the light canopy Turn it clockwise until you have the desired airflow...

Страница 25: ...also be performed at installation relocation and at prescribed intervals as specified by an industrial hygienist safety officer or other qualified person NSF International details the steps for perfo...

Страница 26: ...valve after proper decontamination To wash the drain pan under the work surface lift up and surface decontaminate the solid work surface the perforated grill and remove the work surface supports This...

Страница 27: ...r Once removed the filter should immediately be sealed in a chemical hazard bag and then disposed of safely in accordance with environmental regulations After filter replacement has been completed the...

Страница 28: ...remove the filter by sliding it towards you NOTE once removed the filters should immediately be sealed in a chemical hazard bag appropriately marked for the type of hazard and then disposed of safely...

Страница 29: ...ffuser is installed in the work area under the downflow HEPA filter If your cabinet is connected to an in house exhaust make sure that there is adequate exhaust suction and the system is not producing...

Страница 30: ...The outlets are also protected by a self resetting thermal circuit breaker A qualified electrician using the wiring diagram in the Appendix of this manual can check this breaker If the ultraviolet lig...

Страница 31: ...f a break If the air magnehelic gauge has high or low readings A higher reading than 0 15 to 0 17 inches water column could be an indication of the following Blocked or partially obstructed perforated...

Страница 32: ...lectrical components 9 Disconnect the cord assembly at the top of the unit that goes up from the electrical board assembly 10 Disconnect plug on left side 11 Lift off the whole electrical board assemb...

Страница 33: ...st plenum enough to allow the clamping hardware to angle out away from the slotted retaining brackets 26 Remove the nuts from each petcock plumbing connection 27 Repeat the process on the other side 2...

Страница 34: ...APPENDIX...

Страница 35: ...HT 87 3 8 MIN HEIGHT ONE PLUGGED PENETRATION MAX HEIGHT 91 1 4 MIN HEIGHT 36 5 8 MAX 28 1 8 MIN HEIGHT TO BOTTOM OF UNIT INCLUDES 2 1 2 OF LEG LEVELER ADJUSTMENT 2 3 4 27 1 4 WORK SPACE HEIGHT ONE PET...

Страница 36: ...34 SG303 Ladder Schematic...

Страница 37: ...Switch 16651 Mag Gauge 20773 Interface Board 36504 Motor Blower Controller 304227 Relay Solid State 39636 Sash Balance 39554 Side Panel LH 332X602 Side Panel RH 332X601 Standard Outlet 18231 Supply B...

Страница 38: ...onents hardware and tools Step 2 Place stand sides on a flat floor so that the stainless steel legs are setting on the floor Next place the stand back between the sides so that the open part of the ch...

Страница 39: ...ing the cabinet to the stand and moving it to the room Step 5 Rotate the stand assembly to the upright position so that the leg levelers are now resting on the floor Step 6 Lift the cabinet high enoug...

Страница 40: ...her lock washer and hex nut onto each of the bolts Tighten after all 4 nuts are threaded on Step 8 The lift device should be removed if not being used to move cabinet into position within the room If...

Страница 41: ...st below Item No Description Qty 1 Adjustable Foot or Caster Option 4 2 3 8 16 x 1 Hex head bolt 8 3 3 8 Lock washer 8 4 3 8 Flat washer 8 5 7 16 Plug button 8 OR Slide the unit off the pallet 1 To pr...

Страница 42: ...40 illustration below 4 Screw in two of the adjustable feet or casters item 1 Screw item 1 in by hand as tight as you can and remove the blocks...

Страница 43: ...installed casters to avoid cabinet movement then repeat steps 3 and 4 for the other end of the cabinet 6 Only the adjustable foot option is shown above for clarity Raise the Unit to 30 1 8 work surfa...

Страница 44: ...height adjustable foot option To raise the unit to the 36 work surface height caster option 14 Lift one end of the cabinet up a minimum of 7 and block in position 15 Unscrew the bolts two per leg tha...

Страница 45: ...Company Inc will F O B Sanford Maine furnish without charge repairs to or replacement parts or equipment which proved defective in material or workmanship No claim may be made for any incidental or co...