389D000 Rev B 

3

 

 

 

Standard Features 

Cabinet Primary Function 

The SterilGARD

®

 e

3

 is designed as a Class II, Type A2, Biological Safety Cabinet (BSC) as classified by 

NSF/ANSI Standard 49. This Class of BSC may exhaust to the room or through a canopy exhaust connection 
located at the top of the cabinet into an external facility exhaust system.

 

Applicable Standards 

Except as noted elsewhere in this document or in the factory test report, this cabinet meets the requirements of 
and is tested to UL61010-1, 2

nd

 Edition, CAN/CSA-C22.2 No. 61010-1, 2

nd

 Edition, NSF/ANSI Standard 49-

2012 and IEC/EN61326-1(2006), IEC/EN61000-3-2(2000), IEC/EN61000-3-3(2001) providing NSF listing and 
CE Mark (APPROVAL PENDING) 

Cabinet Pressure Plenums 

The cabinet work area is surrounded by negative pressure plenums, all external seals are under negative 
pressure, and all internal positive pressure plenum seals are surrounded by negative pressure plenums. All 
pressure plenums are ruggedly constructed in order to provide strength, durability, air tight seals and resistance 
to deterioration. 

Motor/Blower Capacity, Cabinet Supply Blower 

The SterilGARD

®

 e

3

 cabinet is designed to automatically compensate for an increase in pressure drop across the 

filters without reducing total air flow rate by more than 10%. The air flow capacity of the cabinet is measured 
by the ability to provide a nearly constant volume of air as the filter resistance to airflow increases. 

Manual Air Balance  

The SterilGARD

®

 e

3

 cabinet has a manual internal balancing damper that compensates for down flow and 

exhaust/intake imbalances.  

WARNING 

This damper must only be adjusted by a qualified technician who has had proper training and has the 

proper equipment for checking the air balance inside the cabinet.

 

 

Cabinet Filters 

Both supply and exhaust filter inside the cabinet are scan-tested HEPA filters. They are 99.99% effective on 
removal of most penetrating particles (mpp) of 0.3 micron size. Each filter is leak checked after installation in 
the cabinet and prior to shipment. 
Optional factory installed Supply and exhaust ULPA Filters are available. The proven efficiency of ULPA 
filters is 99.999% for removal of most penetrating particles (mpp) 0.1 to 0.2 microns in size. 

Easy Filter Access 

All filters are front accessible for ease of service. The supply and exhaust filters can be removed from the front 
of the cabinet through an access panel located behind the fascia panel in the front of the cabinet.  

WARNING 

Only qualified technicians should replace filters. 

Work Area Design 

The interior side and rear work area walls are constructed from a single piece of stainless steel. Inside corners 
are constructed with a smooth radius to help prevent buildup of contaminants and facilitate cleaning. 

Front Access High Velocity Air Slots 

High velocity air slots are placed around the perimeter of the cabinet access opening. Slots are located at the 
front of each sidewall at the access opening, and horizontally along the top of the access opening behind the 
view screen. The purpose of the slots is to improve the cabinet integrity by capturing any particulates at these 
critical transitions. 

Содержание SterilGARD e3 Series

Страница 1: ...B i OPERATOR S MANUAL Class II Type A2 Biological Safety Cabinet MODELS SG404 SG504 SG604 SG404 INT SG504 INT SG604 INT SG404 J SG504 J SG604 J Revision Date Description B 11 12 14 Changed Warranty St...

Страница 2: ...ring a contamination free atmosphere In addition to the high quality you expect from all Baker equipment this model has been designed to provide the lab user with the most comfortable work environment...

Страница 3: ...ea Lighting 4 Electronic Ballast 4 Ground Fault Circuit Interrupter 115V only 4 Laboratory services 4 Alarms 4 Timers 4 Padded Bypass Armrest 5 UniPressure Plenum 5 Motor Blower Assembly 5 Cable Port...

Страница 4: ...increment programming 10 1 hour increment programming 10 Cable Ports 10 Cable Ports continued 11 Start up Procedure 11 Working in the Cabinet 12 Using Ancillary Equipment 13 Reacting to Spills 13 Clea...

Страница 5: ...supply air passes through a diffuser but some of the air is directed down the back side of the view screen creating a high velocity air curtain at the front access opening The HEPA filtered supply air...

Страница 6: ...ILTER POSITIVE PRESSURE PLENUM SUPPLY HEPA FILTER REMOVABLE WORK SURFACE AND SUPPORTS FLUORESCENT LAMP LIGHT CANOPY SLIDING VIEWSCREEN 10 DEGREE SLOPE SUPPLY DIFFUSER FILTER PROTECTOR AUXILIARY OUTLET...

Страница 7: ...cabinet has a manual internal balancing damper that compensates for down flow and exhaust intake imbalances WARNING This damper must only be adjusted by a qualified technician who has had proper train...

Страница 8: ...The SterilGARD e3 standard view screen access opening is 8 203 mm Other optional factory configurations available with NSF listing are 10 254 mm and 12 305 mm access openings Work Area Lighting The w...

Страница 9: ...the recorded value in the most recent certification report However if drastic change is noted this should be cause for investigating the reason This device is not intended to be used for air flow set...

Страница 10: ...rnia OSHPD pre approval FlexAIR Canopy Exhaust Connection CEC The FlexAIR canopy exhaust connection with integrated alarm provides the cabinet with a safe exhaust connection to building exhaust system...

Страница 11: ...BMBL 5th edition published by the U S Department of Health Human Services as HHS Publication No CDC 21 1112 as an advisory document for safe work practices Baker biosafety cabinets are designed for c...

Страница 12: ...e device It is recommended that the GFCI device be tested periodically to assure safe operation Ultraviolet UV Light Optional WARNING UV light is hazardous DO NOT defeat the interlock Eyes and skin sh...

Страница 13: ...le Proximity Sensor Fault Alarm The Double Proximity Sensor Fault alarm occurs when both Sash Position proximity sensors are activated simultaneously This is a condition that should not be able to occ...

Страница 14: ...about an additional 3 seconds you will hear a longer beep This indicates that you have turned the timer function ON are in the programming mode for the device and have programmed it to turn OFF in 1...

Страница 15: ...o they cannot operate simultaneously see also step 2 5 Wipe down the interior area of the cabinet with surface disinfectant NOTE Some disinfectants such as bleach or iodine may corrode or stain the st...

Страница 16: ...s in and out of the cabinet through the air barrier Room airflow can significantly affect cabinet operation Opening and closing doors in the laboratory can cause air disturbances which might interfere...

Страница 17: ...ol or steel pads when cleaning stainless steel Dirt deposits on stainless steel dust dirt and finger marks can usually be removed using warm water with or without detergent If this does not remove the...

Страница 18: ...riate agent prior to maintenance service or repairs in any contaminated area of the cabinet The National Institute of Health National Cancer Institute and the Centers for Disease Control have all reco...

Страница 19: ...shorten the filter s life The High Efficiency Particulate Air HEPA filter is one of the essential components of a clean air cabinet It is the shield which stands between the product and the environmen...

Страница 20: ...he Baker Company Inc will F O B Sanford Maine furnish without charge repairs to or replacement parts or equipment which proved defective in material or workmanship No claim may be made for any inciden...

Отзывы: