background image

 
 

389D000 Rev B 

1

 

 

 

I – FUNCTION OF THE SterilGARD

 e

3

 

 

The SterilGARD

®

 e

is a Class II, Type A2, Biological Safety Cabinet (BSC) as classified by  

NSF/ANSI Standard 49. This Class of BSC may operate with or without an external facility exhaust system. 
The cabinet relies on the flow of air to provide personnel, product and environmental protection.  Personnel 
protection (containment) is provided by a minimum intake air velocity at the front access opening.  Product 
protection is provided by the HEPA filtered supply air in the cabinet work area.  Environmental protection is 
provided by the exhaust HEPA filter. 

 

 
Room air enters the front access opening of the cabinet at a minimum of 100 FPM [0.508m/sec] then enters the 
front work surface perforation. Most of the HEPA filtered supply air passes through a diffuser but some of the 
air is directed down the back side of the view screen creating a high velocity air curtain at the front access 
opening. The HEPA filtered supply air in the work area splits at the work surface. Some of the air enters the 
rear work area perforation while the remainder of the air enters the front work surface perforation. The air is 
pulled through the drain pan area, up the rear and side wall plenums, to the cabinet blower. The blower pushes 
the air into the positive pressure plenum. At that point most of the air is pushed through the supply HEPA filter 
while the remainder is exhausted out the exhaust HEPA filter and through a perforated filter protector at the top 
of the cabinet. [Reference Figure1] 
 

      

 

 

Figure 1: Airflow design 

EXHAUST 

INTAKE 

(ROOM AIR) 

WORK SURFACE 

SUPPLY HEPA 

FILTER

POSITIVE PRESSURE 

PLENUM 

= ROOM AIR 
= CONTAMINATED AIR 
= HEPA FILTERED AIR 

EXHAUST HEPA 

FILTER 

DRAIN PAN AREA 

MOMENTUM AIR 

CURTAIN 

Содержание SterilGARD e3 Series

Страница 1: ...B i OPERATOR S MANUAL Class II Type A2 Biological Safety Cabinet MODELS SG404 SG504 SG604 SG404 INT SG504 INT SG604 INT SG404 J SG504 J SG604 J Revision Date Description B 11 12 14 Changed Warranty St...

Страница 2: ...ring a contamination free atmosphere In addition to the high quality you expect from all Baker equipment this model has been designed to provide the lab user with the most comfortable work environment...

Страница 3: ...ea Lighting 4 Electronic Ballast 4 Ground Fault Circuit Interrupter 115V only 4 Laboratory services 4 Alarms 4 Timers 4 Padded Bypass Armrest 5 UniPressure Plenum 5 Motor Blower Assembly 5 Cable Port...

Страница 4: ...increment programming 10 1 hour increment programming 10 Cable Ports 10 Cable Ports continued 11 Start up Procedure 11 Working in the Cabinet 12 Using Ancillary Equipment 13 Reacting to Spills 13 Clea...

Страница 5: ...supply air passes through a diffuser but some of the air is directed down the back side of the view screen creating a high velocity air curtain at the front access opening The HEPA filtered supply air...

Страница 6: ...ILTER POSITIVE PRESSURE PLENUM SUPPLY HEPA FILTER REMOVABLE WORK SURFACE AND SUPPORTS FLUORESCENT LAMP LIGHT CANOPY SLIDING VIEWSCREEN 10 DEGREE SLOPE SUPPLY DIFFUSER FILTER PROTECTOR AUXILIARY OUTLET...

Страница 7: ...cabinet has a manual internal balancing damper that compensates for down flow and exhaust intake imbalances WARNING This damper must only be adjusted by a qualified technician who has had proper train...

Страница 8: ...The SterilGARD e3 standard view screen access opening is 8 203 mm Other optional factory configurations available with NSF listing are 10 254 mm and 12 305 mm access openings Work Area Lighting The w...

Страница 9: ...the recorded value in the most recent certification report However if drastic change is noted this should be cause for investigating the reason This device is not intended to be used for air flow set...

Страница 10: ...rnia OSHPD pre approval FlexAIR Canopy Exhaust Connection CEC The FlexAIR canopy exhaust connection with integrated alarm provides the cabinet with a safe exhaust connection to building exhaust system...

Страница 11: ...BMBL 5th edition published by the U S Department of Health Human Services as HHS Publication No CDC 21 1112 as an advisory document for safe work practices Baker biosafety cabinets are designed for c...

Страница 12: ...e device It is recommended that the GFCI device be tested periodically to assure safe operation Ultraviolet UV Light Optional WARNING UV light is hazardous DO NOT defeat the interlock Eyes and skin sh...

Страница 13: ...le Proximity Sensor Fault Alarm The Double Proximity Sensor Fault alarm occurs when both Sash Position proximity sensors are activated simultaneously This is a condition that should not be able to occ...

Страница 14: ...about an additional 3 seconds you will hear a longer beep This indicates that you have turned the timer function ON are in the programming mode for the device and have programmed it to turn OFF in 1...

Страница 15: ...o they cannot operate simultaneously see also step 2 5 Wipe down the interior area of the cabinet with surface disinfectant NOTE Some disinfectants such as bleach or iodine may corrode or stain the st...

Страница 16: ...s in and out of the cabinet through the air barrier Room airflow can significantly affect cabinet operation Opening and closing doors in the laboratory can cause air disturbances which might interfere...

Страница 17: ...ol or steel pads when cleaning stainless steel Dirt deposits on stainless steel dust dirt and finger marks can usually be removed using warm water with or without detergent If this does not remove the...

Страница 18: ...riate agent prior to maintenance service or repairs in any contaminated area of the cabinet The National Institute of Health National Cancer Institute and the Centers for Disease Control have all reco...

Страница 19: ...shorten the filter s life The High Efficiency Particulate Air HEPA filter is one of the essential components of a clean air cabinet It is the shield which stands between the product and the environmen...

Страница 20: ...he Baker Company Inc will F O B Sanford Maine furnish without charge repairs to or replacement parts or equipment which proved defective in material or workmanship No claim may be made for any inciden...

Отзывы: