AWA Troubleshooting 

 
 

Troubleshooting 

 

Please review this section before you call Baker Instrument Company or 

return the unit. 

Self Help and Diagnostics 

 
Problems in testing often crop up.  If you are experiencing a problem and believe the 
problem might be with the Baker Instrument Company Advanced Winding Analyzer, 
please take the following steps before calling or returning the unit. 
 
By performing these procedures and having the requested information available, Baker 
Instrument Company’s Service or Applications Departments will be able to better 
analyze your situation and provide the appropriate response.  Either department may be 
reached toll-free at 800-752-8272 or 970-282-1200 for assistance. 

Step #1: Basic Information 

 
Take down all basic instrument information, including the following: 
 

Model #

 

 

 

 

 

example: AWAIII 

Serial # 

 

 

 

 

example: 121 

Product #

 

 

 

 

example: 99-A312-00E 

 

Note:

  All information above is located on the rear panel product label.  If the tester has 

special options installed that you are aware of, please note these.  Any and all information 
you know or can derive would be of help!  A great tool would be a printout or sketch of 
the waveforms displayed on the tester. 

Step #2: Applications or Service Problem? 

 
Generally, if a problem is noted 

Only

 when testing a specific motor/generator or other 

coil type, then Applications would be involved.  See 

Applications: What to do First!

  

Please call Baker Instrument Company, Sales Department for Applications assistance.

Содержание AWA Series

Страница 1: ...may be reached toll free at 800 752 8272 or 970 282 1200 for assistance Step 1 Basic Information Take down all basic instrument information including the following Model example AWAIII Serial example...

Страница 2: ...the waveforms displayed on the tester If you have a FAX available send a draft to 970 282 1010 attn Applications Common Application Problems There are a few common application related problems which m...

Страница 3: ...on the next page The test coil probably has too high an impedance to get a good working pattern The coil may be very high in resistance and turns count The inability to surge test this coil or group...

Страница 4: ...tion Disconnect the test leads from the motor and isolate the tester from any grounded surface Reduce the Output to minimum and attempt a Hipot test with an open lead condition Your display should ind...

Страница 5: ...hese questions the unit is probably operational and indicating you have an open AC line ground connection In the case of numbers 1 through 3 above use an outlet tester to assure proper wiring connecti...

Страница 6: ...wo party operation Always know what test is being performed and WHEN Never attempt a Burn Out of a detected fault with the tester Warranty Return Please review the Warranty Notes and Shipment sections...

Страница 7: ...____________ Phone Number ___________________ Fax __________________________ From the Name Plate on the back of the Tester Baker Product Number ___________________________________________ Model Number...

Страница 8: ...VGA monitor is plugged in 10 AWA never properly concludes a PI test always stops after 5 minutes 11 AWA tester gives a HiPot rip at the activation of every meg ohm HiPot PI test after transport to a...

Страница 9: ...80 123 FAST A D board they will not properly function 25 AWA no longer works after installing a newer version of Analyzer software 26 AWA does not properly save results reporting Unable to update temp...

Страница 10: ...12mm black hex head screws that secure the bottom chassis to the left and right side case L brackets Two are located to the left of the CRT shield two are located to the right of the high voltage tran...

Страница 11: ...successful it will take over the last know good configuration and replace it It is not a tool designed to remove software just a tool designed to help bail out an improper installation b Do not instal...

Страница 12: ...k card is installed a Windows NT is NOT a plug n play capable operating system it may not properly accept a new circuit card without a rebuild of the operating system Check the Microsoft website http...

Страница 13: ...m 80 MB to 30 GB at the time of this writing AWA testers manufactured since 2000 have included vibration dampers on the hard disk These dampers can be retrofitted to ANY AWA manufactured since 1996 b...

Страница 14: ...een the problem is related to the LCD display 9 AWA has a black gray screen with very faint text seen and VGA output appeasrs normal when an external VGA monitor is plugged in a All AWA testers have w...

Страница 15: ...will not apply high voltages without a safety ground 13 AWA gives unbalanced resistance results such as 188 257 189 Ohms on a known good three phase motor a Occasionally it is possible for the auto ra...

Страница 16: ...tests appear to work normally a In AWA Series II tester it was possible for the 80 125 board to be unseated after rough handling Remove the top lid and verify the presence of the 80 125 board in it s...

Страница 17: ...own or damaged U10 IC on the 80 100 AD IO board b Blown or damaged 80 124 HVDC board 17 AWA reaches an arbitrary test voltage for the surge test then stops a Blown or damaged U10 IC on the 80 100 AD I...

Страница 18: ...ms across center to shield when disconnected from the Fast A D board c AD IO control line INT EXT non functional on the 80 100 AD IO board d Connector DB15 J5 knocked loose from the front of the 80 12...

Страница 19: ...and the circuitry responsible for initiating a test Replace the entire assembly or the following I C s b U31 32 Mux08 multi plexers c U10 DAC d U22 AMP01 e U11 SMP11GY f U6 ADC912 g The ADC clock cry...

Страница 20: ...application Don t just hit the circuit breaker You may damage or destroy Windows NT the AWA operating system b This message is almost always related to the 80 100 AD IO board and the circuitry respons...

Страница 21: ...lling newer AWA software to the AWA be sure to select the option for Custom install and then ANALYZER SOFWARE during the install shield process If the user installs the DESKTOP software during a TYPIC...

Страница 22: ...navigate to this file open it edit the following entries and save it b TestSetupPSWRD baker enter what the new password c Password bakerpas enter a new password d Note make sure the file is not set t...

Страница 23: ...e 80 129 stack board If an AWA unit is being serviced be sure to replace this board with the 80 129 002 stack board b Remove the G test lead from the motor switch gear and repeat the surge test If the...

Страница 24: ...5 came from the factory with Micro Solutions Backpack CD R Read Only CDROM This included the BP32DRV sys driver When read write CD ROMs became available the program wrote new drivers to the disk calle...

Страница 25: ...suppressors on the AC line power entry module Remove them replace with MOV s rated for 330VAC b AWA testers were supplied with PATRIOT and SOLA HEVI DUTY un interruptible power supplies between seria...