background image

Masoneilan SVI FF Bench Quick Start Manual 

=

|

 

11

© 2019 Baker Hughes, a GE company, LLC. All rights reserved.

3. Performing Basic Configuration

Figure 9 shows the SVI FF DTM with some key areas highlighted. For more detail, refer to the 
online help.

Figure 9 SVI FF DTM

1. Select 

Quick Start Configuration 

and the 

Quick Start Configuration

 tab appears. Set, as 

required:

These steps under most conditions setup the positioner properly.

DTM help icon

Quick Start Configuration

 tab

Device help icon

Calibration

 tabs in the directory tree

Air Action

Remote Sensor

Control Tuning

Characterization Type

Device Address and Device Tag

Download the Settings

Once all procedures are complete, you must put the SVI FF into 
Normal mode before returning to operations. Do this by:

Choosing Normal from the local display

or using the SVI FF DTM and on the Positioner State tab, clicking the 

Details

 button and setting the state back to Normal

or by setting the parameter 86.APP_MODE to Normal.

1 2

3

4

1

3

2

4

NOTE

Содержание Masoneilan SVI FF

Страница 1: ...Masoneilan SVI FF Advanced Performance Digital Positioner Bench Quick Start Rev E Instruction Manual BHGE Data Classification Public...

Страница 2: ...AL EQUIPMENT IN POTENTIALLY HAZARDOUS ENVIRONMENTS THEREFORE THESE INSTRUCTIONS SHOULD BE INTERPRETED AND APPLIED IN CONJUNCTION WITH THE SAFETY RULES AND REGULATIONS APPLICABLE AT THE SITE AND THE PA...

Страница 3: ...This procedure consists of the following major steps 1 Physical Setup on page 4 2 Software Install and Configuration on page 5 3 Performing Basic Configuration on page 11 4 Optional Find Stops and Ope...

Страница 4: ...486 Connect the leads to the power hub any terminal as follows red lead to green to S black lead to c Plugging the USB plug into USB 8486 into the laptop PC Figure 1 Bench Setup Equipment 4 Mount the...

Страница 5: ...on upgrade via the NI website If NI FBus does not start various tools exist including NI Interface Configuration Utility and NI Dialog utility which are useful in analyzing communications issues ValVu...

Страница 6: ...the advanced features of both ValVue and the SVI FF DTM You then have an additional 30 day period with just standard features after which you must register to continue using the product We strongly e...

Страница 7: ...ppears Figure 3 DTM Library Management 5 Click Update and once the DTM Updates list refreshes select the GE NI FBUS FF DTM and the SVI FF and click Add Sel 6 Click OK to close DTM Library Management d...

Страница 8: ...FF H1 Interface to Network View Figure 5 Figure 5 Network View 9 Right click on the FF H1 Interface and click Add New Device and Figure 6 appears Figure 6 Add New Device 10 Connect the SVI FF to the...

Страница 9: ...the SVI FF DTM See blue box above SVI FF DTM has Address 17 14 Locate the Address field blue box click in it change the address to match the LiveList Address shown in Figure 7 marked by the red box a...

Страница 10: ...og does not appear click Upload All Parameters icon see red box in Figure 8 If you have updated the FF DTM software a dialog appears when you open ValVue 3 Click Yes and the library is updated No and...

Страница 11: ...etup the positioner properly DTM help icon Quick Start Configuration tab Device help icon Calibration tabs in the directory tree Air Action Remote Sensor Control Tuning Characterization Type Device Ad...

Страница 12: ...totune on page 14 Not successful read the Find Stops Log Note the reasons for failure If magnets are out of range then a Click Find Stops tab b Adjust the magnet noting Raw Sensor Value and picture of...

Страница 13: ...e 3 8 required travel 100 4 8 full mechanical travel 79 stop adjustment A sample procedure is 1 Click the Find Stops tab and entering 79 in Open Stop Adjustment field 2 Click the Download Selected Tab...

Страница 14: ...art Configuration Tab for Autotune to run Figure 12 on page 14 For a successful Auto Tune process the supply pressure must be at least 5 psi 34 5 kPa above the spring range WARNING This procedure move...

Страница 15: ...ct an Auto Tune Aggressiveness if necessary 3 Enter an Auto Tune Supply Pressure standard unit only 4 Click 5 When finished ensure the Transducer block is returned to Auto Appendix A Suggested NI Hard...

Страница 16: ...nd enter ValVue in the search field red arrow in Figure 15 This example downloads just ValVue3 separately there is a download that contains all the DTMs as well Figure 15 Download Center Search for Va...

Страница 17: ...If you have a previous install of ValVue3 you are prompted to uninstall first and then you must run the installer again to finish the upgrade NOTE If you are upgrading from ValVue 2 x you must update...

Страница 18: ...OK and it saves to the Windows Downloads folder NOTE For fastest installation save the download file to your laptop PC Don t install from the website 4 Open Windows Explorer and click the Windows Dow...

Страница 19: ...Versions Feature Sub feature Advanced Standard Device States Positioner State X X Positioner Alert Log X X Trend and Position Setup X Device State X X Configuration Control Configuration X X Extended...

Страница 20: ...This page intentionally left blank...

Страница 21: ......

Страница 22: ...111 Moscow Phone 7 495 585 1276 Fax 88 7 495 585 1279 Houston Texas Phone 1 281 671 1640 Fax 88 1 281 671 1735 INDIA Mumbai Phone 91 22 8354790 Fax 88 91 22 8354791 SAUDI ARABIA Phone 966 3 341 0278...

Отзывы: