® 2017 Baker Hughes, a GE company. All rights reserved.
Masoneilan 12400 Series Transmitter/Controller Instruction Manual | 51
9.3. REMOVAL OF 12400 CASE AND TORQUE TUBE SUB-ASSEMBLY (SEE FIGURES 22 & 23)
a.
Switch off power supply.
b.
For instruments with a displacer chamber, close isolation valves, and purge the chamber.
c.
Remove the upper flange (146) and blind flange (144).
d.
Lower the torque arm (135) and unhook the displacer (130). A hook-shaped 3 mm steel wire facilitates unhooking and
holding of the displacer. The wire may be inserted through the clevis hole.
e.
Remove the two torque arm screws (133) and remove the torque arm (135) from the chamber.
f.
Remove the displacer from its chamber (131) or tank.
g.
Be sure that requirements for instruments installed in explosive areas are strictly followed. Remove screw (106) from
connection compartment and unscrew cover (104). Disconnect supply electrical wires and other equipment from terminals
(90).
h.
Remove nuts (142) holding the torque tube subassembly and slide the subassembly out of the mechanism chamber.
PARTS REFERENCE
130
Displacer
143
Gasket
131
Displacer chamber
144
Blind flange
132
Knife of torque tube
145
Stud
133
Torque arm screw
146
Upper flange
134
Torque arm screw block
147
Upper flange stud
135
Torque arm
148
Upper flange stud nut
137
Torque tube housing S/A
149
Upper flange gasket
142
Stud nut
CAUTION:
To remove the displacer, the torque arm must be removed. When the two arm srews (133) are loosened, hold the
torque arm so as not to damage the torque tube (Figure 22).
Figure 22
Figure 23
Содержание Masoneilan 12400 Series
Страница 46: ...44 BHGE 2017 Baker Hughes a GE company All rights reserved...
Страница 62: ...60 BHGE 2017 Baker Hughes a GE company All rights reserved APPENDICES...
Страница 68: ...66 BHGE 2017 Baker Hughes a GE company All rights reserved APPENDIX C see following pages...
Страница 72: ...70 BHGE 2017 Baker Hughes a GE company All rights reserved APPENDIX D see following pages...