Copyright 2022 Baker Hughes Company.
English–PTC Temperature Calibrators | 49
13.7 Heating and Cooling Times
Note:
The heating and cooling times are dependent on the parameters of the test task and the
ambient conditions.
The following times are guide values only. They are measured at a room temperature of 23
°C
(73.4
°F) and do not include transient effects.
Measuring range
-
10 to 100
mV
Accuracy:
- Cold junction
±
0.3
°C (±
0.54
°F)
- Type K
±
0.08
°C (±
0.14
°F)
- Type J
±
0.07
°C (±
0.13
°F)
- Type N
±
0.13
°C (±
0.23
°F)
- Type E
±
0.06
°C (±
0.11
°F)
- Type T
±
0.09
°C (±
0.16
°F)
- Type R
±
0.78
°C (±
1.40
°F)
- Type S
±
0.73
°C (±
1.31
°F)
Current Measurement:
Number of channels
1
Connection
4
mm safety socket
Measuring range
0 to 24
mA
Accuracy
0.01
% full-scale
Voltage Measurement:
Number of channels
1
Connection
4
mm safety socket
Measuring range
0 to 12 VDC
Accuracy
0.01
% full-scale
Switch Test:
Number of channels
2
Transmitter Supply:
Output current
24
mA (maximum)
Output voltage
24
VDC
Specification