6-60
CL 5
6.2 Setting Communication
Speed (Baud Rate) and
Connecting to a PC
Using the proper cable, the instrument’s I/O port can
be connected to a PC. Before connecting to a PC do
the following:
Step 1:
Press to display the CONFIGURATION
menu
Step 2:
Select the COMM (communication) control
Step 3:
Press to activate the control, then or
to adjust the selected baud rate to match that of the
connected device. Data transfer will not occur if the
selected baud rate does not match the device.
Step 4:
Press when the desired value is selected.
6.3 Remote Commands
The CL 5 can receive coded instructions from a
personal computer or terminal connected to the
RS-232 I/O port. Queries, key presses and adjusting
I/O Features
Setting Communication Speed and Type
instrument settings can be done remotely by using
a user written program or a commercially available
serial communications program such as Windows™
HyperTerminal. After starting and configuring
the program, commands are inputted using the
computer keyboard.
Two types of command structures are possible:
• To request the status or value of a function the
following sequence is used:
<ESC><COMMAND><RETURN>
• To execute a keypad operation or adjust a setting
the following sequence is used:
<ESC><COMMAND><SPACE><VALUE><RETURN>
Examples:
<ESC><ID><RETURN>
The CL 5 will now return the model of the instrument.
<ESC><7K><SPACE><8><RETURN>
The CL 5 will respond as if the user has pressed the
mode button.
Содержание 022-508-909
Страница 1: ...Krautkr mer CL 5 Operating Manual 021 002 296 rev 04...
Страница 2: ...CL 5 1 2 General Information 1...
Страница 10: ...CL 5 2 10 Understanding the Keypad Menu System and Displays 2...
Страница 20: ...CL 5 3 20 Setting Up the CL 5 3...
Страница 24: ...3 24 CL 5 Configuring the Instrument s Display Setting Up the CL 5 FIGURE 3 2 Selecting a Setup File...
Страница 26: ...3 26 CL 5 Configuring the Instrument s Display Setting Up the CL 5 FIGURE 3 3 Changing Parameter Settings...
Страница 35: ...CL 5 3 35 Setting Up the CL 5 Locking and Unlocking Instrument Controls THIS PAGE LEFT INTENTIONALLY BLANK...
Страница 36: ...CL 5 4 36 Measuring Thickness 4...
Страница 49: ...CL 5 5 49 Using the Optional Data Recorder 5...
Страница 56: ...5 56 CL 5 Printing a Report Using the Optional Data Recorder THIS PAGE LEFT INTENTIONALLY BLANK...
Страница 57: ...CL 5 6 57 I O Features 6...
Страница 59: ...CL 5 6 59 Transmitting Thickness Data Through an External Device I O Features...
Страница 69: ...CL 5 7 69 Specifications 7...
Страница 76: ...7 76 CL 5 Specifications CL 5 Probe Transducer Specifications THIS PAGE WAS INTENTIONALLY LEFT BLANK...
Страница 77: ...CL 5 8 77 Maintenance 8...
Страница 79: ...CL 5 9 79 Appendix 9...
Страница 84: ...CL 5 10 84 Index 10...