Operating Manual BaehrTec S2000
Page 30/52
English– V101 – Status 07/2015
•
Handpiece motor speed:
6,000 rpm
•
Handpiece motor direction: right
•
Spray level:
OFF
Please note that the selected adjustments are saved if you keep the
memory button pressed for longer than 2 seconds. This means that
settings saved to this memory button will be lost.
In order to recall settings that were saved on a memory button, briefly press the desired
memory button. The saved settings are then applied. This will be indicated to you on the
display by showing the corresponding memory button.
Please note that when briefly pressing the memory buttons, you will only
hear a sound once you remove your finger from the button.
The indication of the memory button will only appear on the display if you
select the settings via the memory button. If you (accidentally) make
settings that are saved to a memory button, you will not be shown on the
display that these settings are already stored on a memory button.
4.8
Getting started
1.
Make sure that there is enough spray solution in your BaehrTec S2000. To check,
please remove the tank lid and fill the tank up to just below the tank rim of the liquid
tank with an
approved spray solution (Baehr spray solution for Baehr spray
technology devices, Art-No. 22333).
Now close the tank lid tight by screwing it on.
Never use liquids that are not approved for the device (e.g. tap water or
self-mixed solutions) since these could cause calcification or algae
deposits on the fine spray lines. This would destroy the pumps and the
water cooling system ( no guarantee on all spray-related units and
components).
On no account should you use emulsions of creams you have mixed
yourself for the spray function since these could damage the spray
function and the guarantee will no longer apply.
Please only use spray liquids that have been approved for the spray
device.
2.
Please ensure that the device has a secure footing, does not fall down and cannot be
torn down, and that the extracted air can escape easily. Please never cover the
ventilation holes. Switch the main switch
(3)
off.
3.
Connect the supplied non-heating connection cable
(57)
with the non-heating
connector
(56)
to the plug
(37).
4.
Plug the safety plug to
(58)
in a proper safety outlet.
5.
Now switch the BaehrTec S2000 on using the main switch
(2)
. After a short LED
check, the LED "Power ON"
(4)
and the LED for the largest instrument (6,000 rpm.)
will light up
(5a)
.
The following things are set after each power on:
o
Handpiece motor direction of rotation = right
o
Handpiece speed = 6,000 rpm (LED illuminated)
o
Handpiece motor = OFF
o
Spray level = 0
o
Spray = OFF
These settings will also be shown to you accordingly on the display. The BaehrTec
S2000 is now ready for use.