Technical Description

3.1 

Intended Use

The s::can micro::station and nano::station are designed for online monitoring of water quality parameters. The 

required components (probes, sensors and operator terminal) are mounted with the required flow cells, mounting 

fittings and pipework on a compact panel. The stations can be used for raw water, drinking water, tap water and 

clean surface water. For waste water and water with high amount of particles a specific micro::station with larger 

pipework and flow cells for waste water is available. 

In all types of applications, the respective acceptable limits, which are provided in the technical specifications in 

the respective s::can manuals, have to be observed. All applications falling outside of these limits, and which are 

not authorised by s::can GmbH in written form, do not fall under the manufacturer’s liability.

The device must only be used for the purpose described in this manual. Use in applications not described in this 

s::can manual, or modification of the device without written agreement from s::can, is not allowed. s::can is not 

liable for claims following from such unauthorised use. In such a case, the risks are the sole responsibility of the 

operator.

3.2 

Functional Principle

All  s::can  stations  are  equipped  with  an  inlet  and  outlet  and  designed  for  continuous  discharge  of  the  water 

through the different flow cells. An inlet strainer will prevent particles from entering the station and a flow restrictor 

ensures correct pressure and discharge. A flow detector alarms when the flow is too low. The water leaves the 

station pressure-free (free atmosphere). With the help of a shut-off valve after the inlet, the flow can be completely 

shut off in case of service activities.

Optional the station can be equipped with an analog pressure sensor and a seperate feed pump.

3.3 Product

The following variants of completely assembled monitoring stations as well as singel components of monitoring 

stations are available. Regarding detailed information of the stations please refer to section 9 (accessories) as 

well as the technical specifications located at the end of this manual. If needed, the monitoring stations can be 

configured customer specific also.

Type

Specification

M-100-MICRO-S

Turbidity, NO

3

, TOC, DOC, pH, EC, Temp, Alarms, mounted on panel with 2 flow 

cells, autobrush, flow detector and con::cube operator terminal; event detection 

software included.

M-100-MICRO-M

Turbidity, NO

3

, TOC, DOC, NH

4

-N, pH, Temp, Alarms, mounted on panel with 2 flow 

cells, autobrush, flow detector and con::cube operator terminal; event detection 

software included.

M-100-MICRO-L

Turbidity, NO

3

, TOC, DOC, pH, EC, Temp, one disinfection parameter, Alarms, 

mounted on panel with 2 flow cells, autobrush, flow detector, and con::cube opera

-

tor terminal; event detection software included.

M-100-MICRO-XL

Turbidity, NO

3

, NO

2

, TOC, DOC, pH, EC, 2 disinfection parameters, Temp., Alarms, 

mounted on panel with 2 flow cells, autobrush, flow detector, and con::cube opera

-

tor terminal; event detection software included.

micro::station / nano::station, 03-2023 Release

www.s-can.at

Copyright © s::can GmbH

7 / 36

Содержание scan micro station

Страница 1: ...Manual micro station nano station March 2023 Release S 36 M...

Страница 2: ......

Страница 3: ...45 ALARM 18 5 2 1 Connection of flow detector F 45 ALARM to con lyte D 320 18 5 2 2 Connection of flow detector F 45 ALARM to con cube 19 5 3 Connection of Pressure Sensor F 500 P 21 5 3 1 Connection...

Страница 4: ...3 Process Connection 30 9 4 Flow Detector 30 9 5 Inlet Strainer 31 9 6 Flow Restrictor 31 9 7 Pressure Sensor 32 9 8 Water Pump 32 9 9 Service Set 33 10 Technical Specifications 34 micro station nano...

Страница 5: ...an by s can This original manual serves as the reference in case discrepancies occur in versions of the manual after translation into third languages This manual and all information and figures contai...

Страница 6: ...self the local legislation and requirements e g regarding electrical connection are observed All s can products are leaving our factory in immaculate technical and safety conditions Inappropriate or n...

Страница 7: ...d a flow restrictor ensures correct pressure and discharge A flow detector alarms when the flow is too low The water leaves the station pressure free free atmosphere With the help of a shut off valve...

Страница 8: ...NANO 8 8 parameters Turbidity TOC UV254 color EC pH 2 disinfection parameters alarm mounted on panel with flow cell autobrush flow detector and con cube operator terminal event detection software incl...

Страница 9: ...s can stations has been optimized Modified dimensions or similar are marked with the note version since 2023 micro station waster water figure on the right micro station nano station 03 2023 Release w...

Страница 10: ...g valve Flow detector digital Process concection Inlet strainer Flow cells for waste water Pressure sensor analog i scan Flow restrictor Flow cell for i scan with autobrush and 2 sensors incl flow res...

Страница 11: ...water in mm For the version until the end of 2022 the total width of the panel is 450 mm The inlet is 135 mm and the outlet 290 mm from the left edge micro station nano station 03 2023 Release www s...

Страница 12: ...Dimensions of micro station waste water in mm micro station nano station 03 2023 Release Copyright s can GmbH www s can at 12 36...

Страница 13: ...mm For the version until the end of 2022 the vertical spacing of the mounting holes is 730 mm The outlet was via the floor outlet of the flow cell micro station nano station 03 2023 Release www s can...

Страница 14: ...eated as household waste It must be disposed and recycled as electronic waste Please assist to keep our environmen 3 5 Scope of Delivery Immediately upon receipt please check the received consignment...

Страница 15: ...ical specifications located at the end of this manual Infrastructure Power supply for controller for operation operational reliability voltage power peak free Correct dimensioning mounting and protect...

Страница 16: ...probes and sensors are to be installed in the flow cells provided which are already mounted on the panel This is done according to the mounting instructions chapter 4 of the respective manuals of the...

Страница 17: ...o the manual of the control ler for operation Check the plausibility of the readings obtained after sufficient running in time at least 15 minutes If necessary calibrate the readings of the i scan to...

Страница 18: ...ght button to switch to the Status display Push the Function button Select the menu Manage sensors and confirm with OK Select the men Add sensor and confirm with OK Select the menu Add digital in and...

Страница 19: ...addition create a cable bridge between 12 VDC red terminal at the bottom of the con cube and the digital input see figure on the right Reconnect the con cube to the power supply and wait until moni t...

Страница 20: ...low Enter the Text for value 0 e g no flow Enter the Text for value 1 e g flow OK Push the button Save to store the configura tion 6 7 8 9 6 7 8 9 micro station nano station 03 2023 Release Copyright...

Страница 21: ...h OK The successful installation is confirmed with ADDED in the top line Push the Back button to return to the menu Sensor configuration Select the menu Configure and confirm with OK Select the entry...

Страница 22: ...om of the con cube with the positiv input e g 1 of the used analog input Analog In with a cable bridge Reconnect the con cube to the power supply and wait until moni tool has started up Click the Serv...

Страница 23: ...number of decimal places Reso lution to 2 Set the Upper limit to 10 and the Lower limit to 0 Check if the Mode is set to 4 20mA and the Error Mode is set to Limit Push the button Save to store the se...

Страница 24: ...word admin1 or your individual user name Click on the icon Outputs and click on the menu entry Digital Outputs in the new window Click on the icon Edit for the 1 digital output digitalOut1 Push the bu...

Страница 25: ...relay e g 45 Relay 1 NO Connect the 12 V power supply of the con cube red terminals at the bottom of the con cube with a cable bridge to the other input e g 46 Relay 1 C of the used digital relay Che...

Страница 26: ...tion itself includes the following steps see also section 8 for typical error patterns Check if any leakage is visible on the flow cells or hose connections screw fittings Check and clean the strainer...

Страница 27: ...e mode on the operator terminal Close the inlet valve Unscrew the transparent cover of the inlet strainer Remove the filter sieve Clean all parts of the inlet strainer with tap water Check the filter...

Страница 28: ...e waiting time after compressed air cleaning Extend discharge 8 2 Return Consignment RMA Return Material Authorization Return consignments of the s can monitoring system or parts of the system shall b...

Страница 29: ...1 US Environment rating IP IP44 IP55 C 31 EU C 31 US 9 2 Inlet Valve Each monitoring station for clean water is permanently equipped with a 2 way ball valve This allows the inlet to the monitoring sta...

Страница 30: ...very F 48 PROCESS F 48 fitting to G 1 Material Nylon PA PVC F 46 PROCESS F 48 PROCESS 9 4 Flow Detector To monitor the minimum flow rate a digital flow detector is installed after the inlet in the mic...

Страница 31: ...at the same time protect the sensors from excessive pressure a flow restrictor is installed in the station after the optical mea surement spectrometer probe or i scan Name Specification Remark Part no...

Страница 32: ...cess connection 1 4 G external thread 9 8 Water Pump Optionally the station can be equipped with a feed pump to ensure sufficient flow If the power is supplied via the con cube the power requirement o...

Страница 33: ...sensor 1 piece Dummy plug for flow cell autobrush incl connecting nut je 2 pieces Safety bracket for sensor 2 pieces Mounting holder for flow cell Replacement brush for autobrush spectromer probe 2 pi...

Страница 34: ...L xx F 508 PANEL per part Required space min 210 mm min 8 27 inch Installation depth Mounting 4 mounting holes 9 mm per panel Material PE HD white PP grey F xxx ECO F 506 PANEL xx F 508 PANEL Material...

Страница 35: ......

Страница 36: ...s can GmbH Brigittagasse 22 24 1200 Vienna Austria Tel 43 0 1 219 73 93 0 Fax 43 0 1 219 73 93 12 office s can at www s can at...

Отзывы: