For further detail (e.g. the detailed connection plan) please refer to the manual of the used devices.
5.2
Connection of flow detector F-45-ALARM
The flow detector of the micro::station / nano::station is pre-installed ex works. In the following sections the elec
-
trical connection and the configuration is explained in detail, in case of retrofitting.
5.2.1 Connection of flow detector F-45-ALARM to con::lyte D-320
■
Disconnect the con::lyte from power supply and
open the cover to the terminal compartment.
■
Connect the two black wires of the flow detector
to terminals 31 and 32 (see figure on the right).
■
Reconnect the con::lyte to the power supply and
wait until the operating software has started up.
■
Use the
Left-
or
Right
button to switch to the
Status display.
■
Push the
Function
button.
■
Select the menu
Manage sensors...
and confirm
with
OK
.
■
Select the menü
Add sensor...
and confirm with
OK
.
■
Select the menu
Add digital in...
and confirm with
OK
.
■
Select the menu
Add flow switch
and confirm with
OK
. The successful instal-
lation is confirmed with
ADDED!
in the top line.
■
Push the
Back
button twice to return to the menu
Manage sensors
.
■
Select the menu
flow switch
and confirm with
OK
.
■
Select the menu
Add parameters...
and confirm with
OK
.
■
Select the menu
Add Flow
and confirm with
OK
. The successful installation is
confirmed with
OK!
in the top line.
■
Push the
Back
button several times to return to the Status display.
■
Now the parameter will be displayed (1 = Flow and 0 = no Flow).
Add digital ADDED!
Add flow switch
Add digital in #1
Add digital in #2
Add para. OK!
Add Flow