Advanced
Software Filter
ACC-Filter
(continued)
Peak Detect
[0 M
3
/Sec
2
]
A
Peak Detect offers a diagnostic view of the acceleration components observed
during flow conditions . This parameter records the “high water mark” of the
measured accelerations component . This value helps to properly configure
the Acceleration Factor parameter . Generally, you set the acceleration factor at
about 75% of the Peak Detect measurement .
To reset the Peak Detect setting, follow these steps from the
Advanced
menu:
1 . Select
Peak Detect
from the
Advanced
menu .
2 . Press [
+
] to reset .
3 . Press
[
E
] to return to the
Advanced
menu .
Software Filter
ZFS-Filter
Description
This software filter operates as a zero-flow stability filter . A specific volume is
defined for a specific time window . If that volume is not measured during the
time window, this volume is ignored and not totalized . The actual flow can be
monitored with the status function . Filter options are:
• Volume USG
• Time
• Status (shows Volume and Time)
Software Filter
IIR-Filter
Description
This software filter operates as an infinite impulse response filter, used to help
suppress erratic flow measurements . Contact Badger Meter Technical Support .
Activation
ON/OFF
Coefficient Min
Numeric entry
Coefficient Max
Numeric entry
Coefficient Status
Numeric entry
Sensitivity
Numeric entry
Hysteresis
m/s
Using the Main Menu Programming Options
Page 52
March 2022