Change factory settings
May 2019
LM_OGP2Adv_BA_99_0045
Page 39
9.
CHANGE FACTORY SETTINGS
Each meter is calibrated at the factory for use with motor oil (see
"Change scale factor" under point 9.3
). The unit of measure is
also selected prior to shipment.
Verify Firmware version
1.
The firmware version and code checksum can be displayed by holding the
TOTAL
and
1
button at the same time.
2.
The last two digits on the lower right are the firmware version.
3.
Meters with version 17 or higher use the procedure outlined in this manual for changing of unit of measure and scale
factor.
9.1
Programming
1.
Push and hold at the same time the
TOTAL
and
AUTO
buttons
2.
PrG will appear on the display
3.
Release the
TOTAL
and
AUTO
buttons
4.
Then press and release in order the
1
,
AUTO
,
10
,
0.1
and
TOTAL
buttons
5.
The current unit of measure will now be flashing indicating that
you have entered the programming mode.
9.2
Change unit for measure
The meter comes with an option to choose 4 different units of measure.
Programming display
1.
The actual unit of measure is flashing when the programming
mode is entered.
2.
Press the
TOTAL
button to toggle between the four options; PT,
QT, GAL, L.
3.
When the desired unit of measure is displayed, press the
RESET
button. The unit of measure icon stops flashing.
4.
If L (litres) is selected, the decimal point begins to flash.
•
You now have the option to change the decimal point to either a period or a comma.
•
To do this, press the
TOTAL
button.
5.
If no scale factor changes are necessary, go to save changes (see
point 9.4
CHANGING THE UNIT OF MEASUREMENT FROM METRIC TO ENGLISH OR FROM ENGLISH TO
METRIC CLEARS THE RESETTABLE TOTAL AND ACCUMULATED TOTAL.
Unit of
measurement
Scale factor digits