Setting Personal Identification Numbers (PINs)
The M1000 offers three levels of access to the individual menus:
A
dministrative,
S
ervice and
U
ser level .
The applicable security level for each menu option is indicated as follows:
A
S
U
Administrative
Service
User
The M1000 security feature allows the option to restrict access to the meter by way of a 6-digit Personal Identification Number
(PIN) . No PINs are set at the factory . The system administrator can set up a single PIN for each of the three different levels
of access:
•
Administration
– allows access to all M1000 menu configuration screens .
•
Service
– allows access to service-level and user-level menu configuration screens .
•
User
– allows access only to user-level menu configuration screens .
1 . To activate password protection, navigate to the
PIN
menu . Set
Control
to
ON
.
2 . Enter Login with the password 000000 .
3 . Return to the
PIN
menu and enter [User], [Service] and [Admin] password .
Once the password protection has been activated, enter your PIN under
Login
.
OTEE:
N
The
Login
screen is available only after
Control
has been set to
ON
.
The lock open symbol appears .
The PIN grants you access to either Administrator, Service or User level with the respective access rights (marked with A, S and
U in this manual) . You can now move to the menu and enter your parameters .
Without a login, you can read all parameters, but cannot change them .
Logging In
To change any parameter in the mag meter, the PIN entered must provide the proper security privilege required by
the parameter .
To enter a PIN, go to the
Login
menu and enter the PIN for the required security level .
Once you are properly logged in, the unlocked icon appears on the meter display .
OTEE:
N
A
PIN Error
message displays if the incorrect PIN is entered .
Logging Out
To log out, enter an unknown PIN, then press
EXIT SAVE
.
M1000 Main Menu Programming Options
Page 20
January 2018
MAG-UM-00379-EN-04