Instruction 0019-9337
9
Oxor® III
3.4.7 Exiting the Setup Screen
Press the
I/O
button at any time to exit the Setup Mode and turn the ana-
lyzer OFF. Note that the last displayed parameter is automatically saved
in memory.
4.0 OPERATION
• Power the analyzer ON
• Wait for the unit to warm up
• Take a gas sample
4.1 Taking a Gas Sample
Turn ON the analyzer by pressing the
I/O
button. Observe that when
power is
fi
rst applied, the software revision level is
fi
rst displayed followed
by a screen that counts down the warm-up period. The warm-up period is
60 seconds.
.
Following warm-up, the O
2
screen appears. This screen directly shows O
2
levels in the range of 0-20.9% and ambient temperatures in the range of
0-100° C (32 to 212° F). If the probe is being used, insert the probe tip into
the area to be sampled.
AMBIENT
Note:
If a sensor error was detected during warm-up, the O
2
Sensor Error
Screen will be displayed. Refer to Section 4.4.
Tip:
An O
2
error will occur when the oxygen sensor's output drops to be-
tween 80 and 90. Consider replacing the oxygen sensor when its ouput level
drops below 100.