MGS-150 Manual
30
6309-9000 Rev 2
6.6. Calibrating Electrochemical (EC) Sensors
There are two adjustments required: zero and span. They are monitored
at 0V and VS using a 0-5V scale. If the sensor range is 0-1000 ppm,
then 5V=1000 ppm.
Step
Calibrating Electrochemical (EC) Sensors
1
Locate Pot VR201 which is used to adjust the zero point.
2
Monitor the output between 0V (negative) and VS (positive).
3
Adjust Pot VR201 to 0V or slightly positive (0.01 V is
acceptable).
4
Locate Pot VR202 which is used to calibrate the range (span)
of the sensor.
5
Monitor the output between 0V (negative) and VS (positive).
6
Expose the sensor to calibration gas and allow to stabilize
(approximately 3 minutes).
7
Adjust pot VR202 to the voltage calculated in section 6.4
(page 29).
6.7. Calibrating Infrared (IR) Sensors
Step
Calibrating Infrared (IR) Sensors
1
Locate Pot VR203 which is used to adjust the zero point.
2
Monitor the output between 0V (negative) and VS (positive).
3
Expose the sensor to nitrogen or zero air until output is stable
(approximately 3 minutes).
4
Adjust Pot VR203 to 0 V or slightly positive (0.01 V is
acceptable).
5
Locate Pot VR202 which is used to calibrate the range (span)
of the sensor.
6
Monitor the output between 0V (negative) and VS (positive).
7
Expose the sensor to calibration gas and allow to stabilize
(approximately 3 minutes).
8
Adjust pot VR202 to the voltage calculated in section 6.4
(page 29).
Содержание MGS-150
Страница 10: ...MGS 150 Manual 10 6309 9000 Rev 2 Figure 2 EC or IR Sensor Components and Wiring...
Страница 11: ...MGS 150 Manual 6309 9000 Rev 2 11 Figure 3 SC Sensor Components and Wiring...
Страница 17: ...MGS 150 Manual 6309 9000 Rev 2 17 Section 3 Housing Dimensions Figure 4 MGS 150 Standard Housing...
Страница 18: ...MGS 150 Manual 18 6309 9000 Rev 2 Figure 5 IP66 Housing with Splashguard...
Страница 19: ...MGS 150 Manual 6309 9000 Rev 2 19 Figure 6 IP66Airflow Duct Mount Housing Figure 7 Exd Housing Units mm...
Страница 33: ...MGS 150 Manual 6309 9000 Rev 2 33...
Страница 34: ...MGS 150 Manual 34 6309 9000 Rev 2...
Страница 35: ...MGS 150 Manual 6309 9000 Rev 2 35...