4 Design and Construction
24
DHV0078 b2 EN
©
b2 electronic
GmbH
Pos.
Name
Description
Earthing connector
Serves as connection point from HVA to earth.
HV output
connector
Serves as connection point from the HVA to the HV test lead.
To connect
Screw the HV test lead into the HV output connector and tighten
Power supply plug
Serves as connection point from the HVA to the power source.
Communication
port
Serves as connection point from the HVA to PC (via RS232) or to a USB device
(via USB flash adapter).
HV switch [on/off]
Activates high voltage.
•
To activate HV output
Press within 10 seconds after START
Navigation knob
Enables user to select options and functions shown on display.
•
To scroll selection up or down
Rotate
•
To enter selection
Click (push in)
Emergency OFF
Activates emergency shutdown.
Operation is only possible when emergency OFF is released.
•
To activate emergency OFF
Press in
Key switch [on/off]
Locks the unit to prevent against unauthorized use.
•
To disable unit
Remove key from the OFF Position
•
To reactivate unit
Replace key and turn to ON Position.
Main switch
[on/off]
Activates the HVA.
This switch is a fuse with integrated magnetic auto-reset 16 A
•
To reset
Turn the main switch OFF and then ON again
Remote control
interlock plug
Provides interlock for the remote switch (i.e. door interlock).
Can be connected to a remote emergency OFF switch, a gate, foot pedal or
dead man switch
External
HV
ON/OFF
Connection for an external HV ON/OFF button
Air vent
Air inlet with filter, for cooling of electronic elements.
Air vent
Air outlet, for cooling of electronic elements.
Display screen
Displays menu, options and status information.
LED red
Indicates HV status.
Red light indicates
High voltage is ON (possible DANGER)
DUT not discharged residual voltage >100 V)
LED green
Indicates HV status.
Green light indicates
High voltage is OFF
Содержание HVA120
Страница 2: ...1 Introduction 2 DHV0078 b2 EN b2 electronic GmbH...
Страница 11: ...2 Safety b2 electronic GmbH DHV0078 b2 EN 11...
Страница 22: ...4 Design and Construction 22 DHV0078 b2 EN b2 electronic GmbH Switches and controls Front panel HVA34...
Страница 33: ...5 Test Procedure b2 electronic GmbH DHV0078 b2 EN 33 Connection diagram vacuum bottle test...
Страница 47: ...5 Test Procedure b2 electronic GmbH DHV0078 b2 EN 47 5 5 Discharge Status...
Страница 62: ...11 Declaration of Conformity 62 DHV0078 b2 EN b2 electronic GmbH 11 Declaration of Conformity HVA34...
Страница 63: ...11 Declaration of Conformity b2 electronic GmbH DHV0078 b2 EN 63 HVA30 5...
Страница 64: ...11 Declaration of Conformity 64 DHV0078 b2 EN b2 electronic GmbH HVA30 7 HVA40 5 HVA54 3...
Страница 65: ...11 Declaration of Conformity b2 electronic GmbH DHV0078 b2 EN 65 HVA54 5...
Страница 66: ...11 Declaration of Conformity 66 DHV0078 b2 EN b2 electronic GmbH HVA60...
Страница 67: ...11 Declaration of Conformity b2 electronic GmbH DHV0078 b2 EN 67 HVA68 2...
Страница 68: ...11 Declaration of Conformity 68 DHV0078 b2 EN b2 electronic GmbH HVA90 HVA94...
Страница 69: ...11 Declaration of Conformity b2 electronic GmbH DHV0078 b2 EN 69 HVA120...