2 Safety 

 

 

10 

DHV0075 b2 EN 

© 

b2 electronic 

gmbh

 

 

Work Safety 

 

 

 

 

 

 

 

 

DANGER 

Electric Shock Hazard 

Never assume that equipment is safe to handle without using the necessary 
safety equipment and earthing procedures. 

 

All procedures must comply with local safety regulations 

 

Always treat exposed connectors and conductors as potential electric shock 
hazards. 

 

DUT must be earthed, de-energized and isolated from all power sources. 

 

All auxiliary electrical apparatus such as switchgear, surge arresters etc. 
must be isolated from the test power source and the DUT. 

 

All cables and connectors must be inspected for damage before use. 
Damaged equipment must not be used. 

 

Earth connections must be made first and removed last! 

 

DUT must be discharged and earthed before disconnecting the test lead. 

 

Avoid testing alone. In the event of an emergency another person’s presence 
may be essential.  

 

 

 

Содержание TanDelta TD120-MC

Страница 1: ...er Manual E N G L I S H DHV0075 Rev06 b2 electronic GmbH Riedstra e 1 6833 Klaus AUSTRIA Tel 43 0 5523 57373 Fax 43 0 5523 57373 5 www b2hv com info b2hv com Subject to alterations errors excepted Ill...

Страница 2: ...1 Introduction 2 DHV0075 b2 EN b2 electronic gmbh...

Страница 3: ...of Contents 1 INTRODUCTION 4 2 SAFETY 8 3 GENERAL DESCRIPTION 14 4 DESIGN AND CONSTRUCTION 23 5 CONNECTION UP TO THE TD TD MC INSTRUMENT 26 6 SOFTWARE TD CONTROLCENTER 29 7 INSTRUMENT CARE 51 8 GLOSS...

Страница 4: ...miliarize themselves with this document in full The following is a guideline indicating the most significant information as a function of the user s responsibilities User Responsibilities Focus TD Ope...

Страница 5: ...ot avoided could result in death or serious injury Caution CAUTION Indicates a hazardous situation which if not avoided could result in minor or moderate injury Notice NOTICE Indicates suggested pract...

Страница 6: ...ary items This warranty and our liability are limited to replacing or repairing defective equipment at our discretion Equipment that is returned to b2 must be packed in original packaging All shipped...

Страница 7: ...o a quick and comprehensive support and product data base The b2 Customer Portal is reserved for b2 electronic customers Submit the registration and find Information about meetings and training sessio...

Страница 8: ...t before using the instrument Make sure you are knowledgeable about the proper application safety potential hazards involved and procedures in using the test instrument The awareness for safety is an...

Страница 9: ...interruption of the grounding connection may create an electric shock hazard The visual signals of the TD are designed to be easily distinguished even under bright sunlight to avoid incorrect readings...

Страница 10: ...hock hazards DUT must be earthed de energized and isolated from all power sources All auxiliary electrical apparatus such as switchgear surge arresters etc must be isolated from the test power source...

Страница 11: ...trical technicians should have access to the test area Other persons must be accompanied by qualified electrical technicians and must be informed of the risks involved NOTICE Equipment Handling DUT mu...

Страница 12: ...eer to now make tan delta testing a routine maintenance test The TD is a battery powered system that is directly connected to the HVA series of VLF test systems Standard C size alkaline or rechargeabl...

Страница 13: ...rator Qualifications TD operators must be qualified electrical technicians Proof of necessary qualifications for working in high voltage domain is mandatory It is highly recommended that operators hav...

Страница 14: ...ms Accuracy 1 of reading Current measurement Resolution 1 x 10 5 Accuracy 1 x 10 4 Load range Standard 500 pF to 10 F Weight 3 kg 5 kg without cable 14 2 kg without cable Dimensions mm 240 long 80 in...

Страница 15: ...to 10 F Weight 11 kg without cable 16 kg without cable 16 kg without cable Dimensions mm 660 long 120 in diameter L 300 x H 790 x W 276 L 300 x H 790 x W 276 Computer includes Bluetooth Delivery incl...

Страница 16: ...it Solid state air insulated design Suitable for use with the HVA VLF test systems Test results are easily stored via Notebook Computer or PocketPC via wireless Bluetooth connection Supplied complete...

Страница 17: ...No Item Description TD60 MC TD90 MC TD120 MC DHV0075 TD Manual b2 1 1 1 GH0522 Earthing Cable Transparent 6 mm2 4 m with 400 A clamp 1 GH0526 Earthing cable transparent 6 mm 10 m with 600 A clamp 1 G...

Страница 18: ...100 red 1 1 1 KES0022 Delfin Clamp 4mm connector red 1 1 1 KES0105 Angle Bracket Connector 14mm 1 1 1 KES0209 MC socket connection for corona shield 1 1 1 KMD0081 Corona Shield 2 parts 2 2 2 KN0009 C...

Страница 19: ...ries please contact b2 Art No Item Description TD30 TD60 TD90 DHV0075 TD Manual b2 1 1 1 KDD0016 USB Pen Flash Drive b2 1 1 1 KDG0010 USB Bluetooth Dongle 1 1 1 KEB0002 Extra lid to BX002 1 1 1 1 KEB0...

Страница 20: ...9 MC socket connection for corona shield 1 KMB0050 TD30 Hook 2 KMD0065 TD M8 30 4mm Adapter 3 3 3 KMD0081 Corona Shield 2 parts 2 2 2 KMD0086 HV plug 50 kV Protection 1 KMD0113 TD60 Rack 1 KMD0182 TD...

Страница 21: ...l 2 2 GH0526 Earthing cable transparent 6mm 10m with 600A clamp 1 GH0548 TD90 Cable 100kV 10m Plug Plug 1 GH0550 TD Connecting cable TD charge 0 70 m 1 GH0551 TD Connecting cable TD charge 1 60 m 1 GH...

Страница 22: ...tion 22 DHV0075 b2 EN b2 electronic gmbh Art No Item Description TD30 TD60 TD90 VKR0005 TD60 Set transport case with Inlays 1 VKR0013 TD30 Set transport case with Inlays 1 VKR0026 TD90 Set transport c...

Страница 23: ...f different connection options to the DUT Unscrew head cap section to remove from the rest of the housing to turn ON OFF the system or to replace the batteries To re install the head cap section caref...

Страница 24: ...er no voltage 1kV has being sensed for 60 minutes or more In this mode the main switch has to be set to 0 off for about 2 minutes and then on again to wake up the system manually or you can reapply vo...

Страница 25: ...ntrolCenter PC Software LED 3 Signal Low TD GREEN A flashing signal here indicates a TD level that is lower than the LOW limit If setup correctly this will normally indicate a good TD result The limit...

Страница 26: ...ual The Ground Cable of the TD TD MC should be connected to the station ground In the case of a cable this can be the concentric neutral ground shield ground wire The cable ground must be connected to...

Страница 27: ...earthing connector S2 Connect Power Supply Cable Connect the power supply cable to the HVA power supply plug S3 Connect HV Test Lead Screw the HV test lead which is connected to the TD into the HVA HV...

Страница 28: ...earthing connector S2 Connect Power Supply Cable Connect the power supply cable to the HVA power supply plug S3 Connect HV Test Lead Screw the HV test lead which is connected to the TD into the HVA HV...

Страница 29: ...n purposes See also the TD System Bluetooth Setup Manual For Bluetooth connection setup using the supplied Bluetooth dongle please refer to Appendix A Description of Elements of the Main Screen System...

Страница 30: ...nd the updated mean values of this block The Change Phase Button enables the user to move a selected block to another Phase or to completely remove it from the Report This is useful if phase test data...

Страница 31: ...e ID if cable as DUT is selected The user can enter a unique identifier for the Device Under Test DUT This Text will also be used by default as the title for this Report on the Summary and Printout St...

Страница 32: ...orrect phase section using the Change Phase button in the block header as a later stage Graphical View Diagram This screen provides a graphical diagram of the Mean TD TD MC measurement data versus the...

Страница 33: ...2 Configure Bluetooth support Initial usage only has do be done once Step 3 A B Setting up the TD Communication Port A With use of External Dongle or B Windows Supported Hardware There are 2 different...

Страница 34: ...ge of approximately 50 feet 15 meters Start Installation Start the installation process by inserting the driver CD supplied The Setup should start automatically if not locate the CD DVD drive on My Co...

Страница 35: ...IB9 describe how to Install the Bluetooth Software Step Procedure Installation Bluetooth Software IB1 Setup Start Driver Setup Start Screen IB2 Accept License Accept License agreement IB3 Choose Dest...

Страница 36: ...allation has started IB6 Accept Driver Press OK to accept this driver IB7 find Bluetooth The driver now needs to have the Bluetooth dongle inserted Plug the dongle into a free USB port The windows Har...

Страница 37: ...Installation Bluetooth Driver Installation of Bluetooth support on your Computer Other information messages may appear Please wait until the Installation completes IB9 Installation Finished Installati...

Страница 38: ...gure Bluetooth Support using external dongle Please follow this instruction to set up a communication port to the TD system if you are using the external dongle with the above mentioned driver not win...

Страница 39: ...ribe the Bluetooth Software configuration Step Procedure Bluetooth Software configuration CB1 Bluetooth Configuration Initial Bluetooth configuration only has to be done once CB2 Computer Title Type i...

Страница 40: ...oth services Check or uncheck those Bluetooth services you want to use CB5 Bluetooth Serial Port Configuration Bluetooth Serial Port Configuration Leave as suggested Take note of the serial port numbe...

Страница 41: ...6 Software TD ControlCenter b2 electronic gmbh DHV0075 b2 EN 41 Step Procedure Bluetooth Software configuration CB7 Bluetooth Configuration Finished Finished Initial Setup of Bluetooth Support...

Страница 42: ...TD system Before connecting reset the TD System by switching the main power switch off and then ON again Make sure the TD system is within 50 foot 15 meters of the connecting computer Since the Bluet...

Страница 43: ...1 Bluetooth Setup Wizzard My Bluetooth Places Select Bluetooth Setup Wizard to connect to TD system VT2 Select Bluetooth Device Select I want to find a specific Bluetooth device to find the TD system...

Страница 44: ...Type in the PIN code welcome and press Initiate Pairing A message No Com port available may be appear Just quit with ok VT5 TD Communication Check the TDCommunication service to use A message to confi...

Страница 45: ...nter b2 electronic gmbh DHV0075 b2 EN 45 Step Procedure Connection with a TD Unit VT7 Communication Port Setup Successful Setup of communication port VT8 TD Unit The TD system is now shown as an icon...

Страница 46: ...System Connection Established Connection Read the information and note the COM Port Number This number is used by the TD ControlCenter to communicate with the TD unit Step Procedure Initial Connection...

Страница 47: ...ur desktop This gives you access to all Bluetooth specific properties and configurations Switch on the TD system Before connecting reset the TD System by switching the main power switch off and on Mak...

Страница 48: ...al Connection with TD Windows KT1 Bluetooth Device Bluetooth Devices press Add to start KT2 Add Bluetooth Device Check Device is switched on and ready to discover KT3 Select Bluetooth Device All avail...

Страница 49: ...y Select Use Passkey from documentation and enter welcome to use as key KT5 Install Drivers Windows connects to the TD System and installs the relevant drivers KT6 Finish de Bluetooth Wizard Read the...

Страница 50: ...th TD Windows KT7 Specify a passkey Select Use Passkey from documentation and enter welcome to use as key KT8 Bluetooth Device is configured The System is now configured You can start working with the...

Страница 51: ...only the gasket ring In the case the ring is brittle you should contact b2 electronic GmbH for a spare one KMK0022 CAUTION Instrument Damage Do not store the TD units outdoors Keep the TD units away f...

Страница 52: ...limits Fault An unplanned occurrence or defect in an item which may result in one or more failures of the item itself or of other associated equipment source IEC Frequency Hz Number of cycles per unit...

Страница 53: ...9 Declaration of Conformity b2 electronic gmbh DHV0075 b2 EN 53 9 Declaration of Conformity The TD units are CE certified and have met the following requirements of the European Council...

Страница 54: ...9 Declaration of Conformity 54 DHV0075 b2 EN b2 electronic gmbh...

Страница 55: ...9 Declaration of Conformity b2 electronic gmbh DHV0075 b2 EN 55...

Отзывы: