3 General Description
©
b2 electronic
gmbh
DHV0075 b2 EN
15
Characteristic
TD90
1
TD90-MC
1
TD120-MC
1
Article Number
SH0214
SH0241
SH0245
Name
TD90 Tan-Delta System
TD90 Tan-Delta System
64kVrms with MC
TD120 Tan Delta System
85kVrms with MC
Input Voltage
2 type „C“ alkaline cells
or NiMH batteries
2 type „D“ alkaline cells or NiMH batteries
Operating
Voltage
Sinusoidal
1 – 64kV rms
1 – 85 kV rms
Frequency
0.1 Hz
0.01 to 0.09 Hz
2
Voltage
measurement
Resolution
0.1 kv rms
Accuracy
1% of reading
Current
measurement
Resolution
1 x 10
-5
Accuracy
±1 x 10
-4
Load range
Standard
500 pF to 10µF
Weight
11 kg (without cable)
16 kg (without cable)
16 kg (without cable)
Dimensions (mm)
660 long 120 in diameter
L 300 x H 790 x W 276
L 300 x H 790 x W 276
Computer includes
Bluetooth
Delivery includes
HV connection cable,
transit case,
BluetoothTM Dongle,
Earth cable, TD Control
Center Software,
Operating manual
transit case, BluetoothTM Dongle, Earth cable, TD
Control Center Software, Operating manual
Standards
EMV: IEC6100-4-2, IEC6100-4-4, EN55077;
Safety: EN60950, EN50191, EN61010-1
Environmental Conditions
Storage: -25°C to +70°C , Operating: -5°C to +45°C
Humidity: max. 80% R.H. (non-condensing)
1
Technical Specifications are subject to change. b2 reserves the right to modify values in accordance with future HVA
development
.
2
in combination with locator set (not in scope of supply)