C100 Series Technical Manual
- 19 -
CALIBRATION linear
Turn on the scale and press
during the self test.
Use the numeric key to enter the password
9999
Press
to confirm
Display will be show
Select Local or Remote scale by pressing
Press
to confirm.
Display will be show
If necessary, press
to select the weighing unit
kg
or
lb.
Press
to confirm.
Display will be sho
Ensure the platform is empty and wait for stable indicator.
Press
to confirm.
Display will be show
Apply the weight on the display (indicated in chart below) wait for stable
indication. Then press
to confirm.
Display will be show
Apply the weight on the display (indicated in chart below) wait for stable
indication. Then press
to confirm.
Model
Unload
Load x
Load xx
C100 / 6lb
0 lb
2 lb
6 lb
C100 / 12lb
0 lb
4 lb
12 lb
C100 / 30lb
0 lb
10 lb
30 lb
C100 / 60lb
0 lb
20 lb
60 lb
After linear calibration is complete the scale will start a self test, remove
the weight during that time and display will return to weighing mode.
Pi n
Pi n
- - - -
tech
Local
remote
tech
unit
unload
Load x
Load xx