Scale Calibration
Calibrating Analog Scale
At the main screen press and hold the BTEK Logo, Once Red enter 2835 and tap the Metro
Button on page 8 of 11. Press scale configuration, then on Scale A or B.
1.
Press the calibration Button or enter software seal to access parameters.
After removing metrological seal, the background will turn to gray and the
shield symbol will disappear. Press A to enter parameters.
2.
Press Calibration Procedures to enter inside the calibration menu.
3.
Chose the calibration type to perform. For example, Zero Full Scale will be a
zero and span calibration. The full-scale weight value will be changed to your
test load value.
4.
Check if the scale is empty and press Calibrate to calibrate the dead load.
5.
Press on the Weight Box (120000 lb in the example), enter the test weight,
load the weight on the platform, and press enter.
6.
Press Calibrate to execute the span calibration.