12
Safety instructions
–
Staff call cable compact
plus
(8718030)
–
Connection lead 12 V (8718020)
–
Short infusion stand (8713135)
–
Accessories according to section 20.1
Note:
Special informations with regard to
EMC are included in the separate instruc-
tion manuals for use for each relevant
accessory. Ensure adequate protection
against free-flow before changing dispos-
able items.
Q
Hydrophobic filters can further reduce
the infusion of microbubbles.
Q
See the corresponding manufacturer
information for possible incompatibili-
ties between the device and medicinal
products.
Q
Use only Luer lock, ENFit or NRFit
feed systems, and use only compatible
combinations of devices, accessories,
wear parts and disposable items.
Q
Connected electrical components must
comply with IEC/EN specifications
(e.g., IEC/DIN EN 60950 for data pro-
cessing equipment). Anyone who con-
nects additional devices is considered
a system configurer, and is therefore
responsible for compliance with sys-
tem standard IEC/DIN EN 60601-1-1.
Q
Where several infusion lines are con-
nected on one single vascular access,
the possibility of the lines exerting a
mutual influence over each other
cannot be excluded.
Note:
The use of untested or incompatible
disposable items can affect the technical
data. PUR lines can’t be used with the
device.
4.1.7
Alarms and staff call
Q
The volume of the device’s acoustic
alarms can be adjusted for the envi-
ronmental conditions. This ensures
that the alarms are clearly audible.
Q
Always monitor the pump alarms.
The use of the accessory cable or staff
call does not adequately replace moni-
toring the alarms.
4.1.8
Accessories and consumables
Q
It is recommended that disposable
items are changed after 24 hours
(see hygiene rules).
Q
Only use pressure-tested disposable
items (min. 2 bar/1500 mmHg).
Q
Only use the device with accessories
and articles that have been approved
for use with the device.
WARNING! The use of accessories,
transducers, power cords and cables
other than those specified, with the
exception of those sold by B. Braun
Melsungen AG as replacement parts
for internal components, may result
in increased emissions or decreased
immunity of the Infusomat
compact
plus
P
.
Recommended equipment, accessories,
transducers and cables for which B. Braun
Melsungen AG claims compliance with
the requirements of the standards in
section 4.3:
–
Perfusor® compact
plus
(8717030)
–
Infusomat compact
plus
(8717050)
–
Station compact
plus
(8717141)
–
Cover compact
plus
(8717145)
–
Data module compact
plus
(8717160)
38932401_Infusomat_compactPlusP_EN_IFU_1020_v5.indd 12
15.10.2020 17:47:25