53
Electromagnetic compatibility
18.3
Recommended safe
distances between
portable and mobile
RF telecommunications
equipment and the
Infusomat compact
plus
The Infusomat compact
plus
is designed for
use in an electromagnetic environment in
which emitted RF disturbances are con-
trolled. Customers or users of the Infusomat
compact
plus
or its components can help
prevent electromagnetic interference by
complying with the minimum distances
between portable and mobile RF telecom-
munications devices (transmitters) and the
Infusomat compact
plus
and its components,
as recommended below in accordance with
the maximum output power of the commu-
nication device.
Note:
The higher value applies at 80 MHz
and 800 MHz.
Note:
For transmitters whose rated power is
not specified in the table above, the distance
can be determined using the equation for
the relevant column. P is the transmitter’s
rated power in W according the manufac-
turer’s specifications.
Note:
A factor of 10/3 is used to calculate
the recommended safe distance of trans-
mitters in the frequency range between
80 MHz and 2.5 GHz, in order to reduce
the probability of a mobile communication
device used unintentionally in the patient
area causing a fault.
Transmitter
rated power
in W
Safe distance according to transmitter frequency m
150 kHz to
80 MHz 1.2√P
80 MHz to
800 MHz 1.2√P
800 MHz to
6 GHz 2.3√P
0.01
0.12
0.12
0.23
0.1
0.38
0.38
0.73
1
1.2
1.2
2.3
10
3.8
3.8
7.27
100
12
12
23
38932401_Infusomat_compactPlusP_EN_IFU_1020_v5.indd 53