38
Cleaning and care
2. Do not spray disinfectants directly on
the pump, use a soft, wet and lint-free
cloth. Always use a new cloth to avoid
cross contamination. Wet all surfaces
sufficiently and observe the required
exposure times according to the
manu facturer’s instructions.
3. Inspect all connectors for residual
moisture and evidence of damage.
In case of damage the device should
remain unplugged until it can be
inspected by a trained technician.
Recommendation:
Use disinfectants manufactured by B. Braun:
Meliseptol® Foam pure, Meliseptol® Wipes
sensitive, Melsitt®, Melsept®, Hexaquart®
or Hexaquart® forte.
Substances from the groups of disinfect-
ants listed below are approved, for normal
cleaning according to the manufacturer’s
instructions:
Group
Active Substance
Alcohols
1-Propanol,
2-Propanol (Isopropanol),
Ethanol
QAC
(Quaternary
ammonium
compounds)
DDAC (Didecyldimethyl-
ammoniumchloride),
BAC (Benzalkonium-
chloride)
Acids
Citric Acid,
Milchsäure,
Essigsäure
Phenols
o-phenylphenol,
p-Chlor-m-cresol
10 Cleaning and care
Q
Device is switched off
Q
Device is unplugged from the mains
Q
Device accessories are disconnected
10.1
Cleaning & Disinfection
Cleaning and disinfection can be carried
out by qualified medical professionals or by
qualified and trained cleaning personnel.
CAUTION! Before disinfecting the pump,
always disconnect the pump from the
patient, switch off the device and discon-
nect from power and other devices (e.g.
staff call).
WARNING! Do not allow liquids or
detergents to come into contact
with accessory ports, electrical con-
nections of the device or any device
openings. Fluid exposure in these
areas may result in the risk of short
circuit, corrosion or breakdown of
sensitive electrical components, and/
or electrical shock.
Before use, the device must be completely
dry.
Handling
1. Make sure to remove any visible
residue from all surfaces prior to dis-
infecting. Clean all exposed surfaces
using a clean, soft, lint-free cloth
dampened with a mild cleaning solu-
tion of warm, soapy water.
38932401_Infusomat_compactPlusP_EN_IFU_1020_v5.indd 38
15.10.2020 17:47:30