Programming the Neurostimulator Stimulation Settings
51
Electrode Configuration
The CP provides the ability to switch between monopolar
and bipolar stimulation modes, to use automatically
generated recommended electrode configurations, and
1
to manually set which electrodes are active and inactive
during stimulation.
Monopolar vs Bipolar Stimulation
2
The
Programming
screen allows the electrode
3
configuration to be set to be “Monopolar” or “Bipolar”
(
①
).
If “Monopolar” is selected, the Neurostimulator is
set as the anode and tined lead electrodes can only be
assigned as cathodes. If “Bipolar” is selected, a tined lead
electrode must be selected as an anode to deliver
stimulation.
Electrode Recommendations
The
Programming
screen displays up to four
recommended electrode configurations
(
②
)
. These
recommendations are intended to provide directional
guidance regarding which electrode configurations to
test. The recommendations are generated based on the
stimulation thresholds and motor and sensory responses
4
recorded during tined lead placement or subsequent
redefining of the thresholds. Ultimately, electrode
configuration selection should be based on patient
comfort and symptom reduction. To set the electrode
configuration to a recommended configuration, press the
image of the recommendation.
Note:
If thresholds are not saved to the Neurostimulator, no electrode recommendations are provided.