Setting up a New Neurostimulator
40
Setting up a New Neurostimulator
There are several steps to set up a new Neurostimulator.
A.
A prompt will appear when a new Neurostimulator is
connected after a lead implant, asking for permission
to transfer threshold data that was recorded during
the tined lead placement procedure. A summary of
1
the threshold data will then be saved to the device
2
for reference during programming sessions. This
3
threshold data summary can be viewed in the
“Threshold Detail” tab of the
Patient Device
screen
(
①
)
.
B.
Input the serial number of the patient’s Remote
Control in the Remote Control ID field
(
②
)
. This pairs
the Remote Control to the Neurostimulator, allowing
the Neurostimulator to receive commands from the
Remote Control.
Note:
The Neurostimulator can only communicate with the
Remote Control that is entered in this field.
C.
Enter the patient identifier associated with this
device into the Patient ID field
(
③
)
.
D.
Enter the date of the Neurostimulator implant
procedure in the Device Implant Date field.
These fields, including the Patient ID and the Remote
Control serial number can be updated at any time as
needed.