Viewing the Status of a Neurostimulator
46
Therapy History tab (
①
)
1
This tab shows the stimulation settings saved to the
Neurostimulator in the previous 4 programming
sessions. For each set of stimulation settings, additional
information is provided including the time period the
settings were used and notes entered during the
programming session.
Threshold Detail tab (
②
)
This tab summarizes the stimulation threshold data
previously captured. For each electrode, the tab shows
the motor response, sensory response, and threshold
stimulation amplitude recorded during the tined lead
implantation procedure or during a programming
session. The color around the threshold amplitude
indicates the qualitative assessment of the electrode
2
placement (Red = “Poor”, Yellow = “OK”, Green =
“Good”).
E0
E1
E2
E3