Reference Section
•
123
Chapter 5
Theory of R
s
Compensation
The MultiClamp 700A uses a dual approach for R
s
compensation, like the Axopatch
200 series of amplifiers. This provides superior correction and stability.
For R
s
compensation to function properly, whole cell compensation must have been
adjusted and the Whole Cell checkbox must be checked. Whole cell compensation
provides estimates of R
s
and C
m
, which together determine the shape of the correction
current that is injected through capacitor C2 (Figure 4.20). Note that this C2 correction
current does not improve the speed of clamping of the cell; rather, it charges the
membrane capacitance as slowly as before but in a way that is invisible to the user,
because it bypasses the feedback resistor in the headstage.
Figure 4.20.
Schematic whole-cell compensation circuit.