Reference Section
•
99
Chapter 5
Input/Output Connections
•
Description of the different connectors on the front and rear panels of the
MultiClamp 700A main unit.
•
See also External Command Inputs, Oscilloscope Triggering, Mode.
Front Panel
Inputs
MODE: This is enabled when the user has checked the Ext checkbox under
Channel 1 or 2 Mode in the MultiClamp Commander. A TTL Low input
at MODE will select I-Clamp; a TTL High (3.5-5 V) input will select V-
Clamp. For example, these inputs can be a TTL Digital Signal controlled
by pCLAMP.
COMMAND: Voltage or current commands to the MultiClamp 700A are
accepted at this input. The External Command Sensitivity is set in the
Gains panel under the Options toolbar button.
Outputs
SCALED: This is intended to be the primary conditioned output of the
MultiClamp 700A. Its identity is selected from the list in the Output
Signals section of the main window of the MultiClamp Commander:
•
Command Potential (V-Clamp) or Command Current (I-Clamp)
•
Membrane Current
•
Membrane Potential
•
100 x AC Membrane Potential (High-passed filtered at 1 Hz, this
special high-gain output is useful for viewing very small extracellular
signals.)
•
Bath Current (available when a Bath headstage is used).