TECHNICAL AND MOTION DOCUMENTATION
___________________________________________________________________________________________________
16
3.
After displaying
F..-2
press
ENTER
. The display shows value of the lower threshold,
and after pressing
key – upper threshold. If the printer is connected, threshold
values are printed.
4.
Press
key to switch to weighing.
Caution:
After the balance turning on both thresholds are set to maximum values. During upper
threshold setting pay attention, so its value is not lower than the lower threshold.
After the balance turning off and on this function remains unchanged.
10.4. Automatic zeroing function (AUt)
The function activation will cause automatic maintaining of zero indications, when the
pan is not loaded, or when zero reading has been achieved by pressing "T/
ON
" key.
Operations sequence:
1.
Press key
MENU
.
2.
When the display shows
F..-AUT
press
ENTER
.
3.
The display will show F..-0 and F..-1.
4.
Press
ENTER
when the display shows
F..-1
. On the left side "AUT" message will be
displayed.
5.
To deactivate automatic tare setting function, perform section 1-3, selecting F..-0.
10.5.
Pieces counting function (F..-LIC)
That function allows for counting the identical details, e.g. screws or transistors,
contained in the weighed batch, and may be used as auxiliary function, operating outside
of computer system.
The measurement takes place in two phases:
- first phase – calculation of single detail weight upon the basis of sample of specified
number of pieces
- second phase – counting the details in the weighed batch.
It is recommended, that the single detail weight is higher than the balance reading
graduation, and the weight of sample, used in the first phase is higher than 100 of
reading graduations.
Operations sequence:
1.
Tare the balance pressing T.
2.
Put the sample of 5, 10, 20, 50, 100, 200, or 500 pieces on the pan (pay attention so the
total load is not lower than 100 reading graduations).
3.
Press
MENU
key to recall functions menu. During displaying
F...-LIC
press
ENTER
.
4.
The display will show F..-0 and F..-1. Press
ENTER
when the display shows
F..-1
.
5.
When next to F.. the number of pieces on the pan will be displayed (- -, 5, 10, 20, 50,
100, 200 or 500) press
ENTER
.