USER MANUAL
21
_____________________________________________________________________________________________________________________
14.3 Pieces counting (PCS)
This function enables to count
identical pieces, e.g. turnbuckles or
buttons.
A measurement is performed in two
phases:
- first phase - single piece weight
calculation on the basis of defined
pieces amount (
5, 10, 20, 50, 100,
200
or
500
pieces),
-
second phase – pieces counting.
It is advised that single piece weight
is not less than one reading unit and
sample weight used in first phase is
bigger than 100 reading units.
To leave function press
MENU
key
and then with
→Τ←
key chose
PCS
and
PCS-0
.
Note:
1. Err-3 communicate signalises that
a sample was not put on the pan.
The same
communicate
appears
if single piece weight is less than one
reading unit (it is possible to count
pieces but measuring error is
bigger).
2. To chose previously used pieces
amount select " _ _ " in first phase
(in case no value was chosen, error
communicate appears
3. During pieces counting
→
T
←
key
function does not change.
4.In scales equipped with LCD
display, weighing unit is visible and
" " sign is replaced with "pcs ".
0
0.00
23.40
PCS
PCS -0
PCS-1
PCS - --
5
5
12
12
T
T
T
T