5
9.
Increase more value key
10.
Reduce more value key
11.
Increase less value key
12.
Reduce less value key
13.
Low power indication
14.
Input state indication
15.
Output state indication
16.
Result value
17.
Voltage V mV indication
18.
Current mA mA% indication
Operation Instructions
DC V measurement
1
Press the power switch
5
, turn on the Calibrator
2
Press the input/output conversion key
, when the state of no input indicator. Make it under
the state of measurement.
3
Press the V MV conversion key
6
, make it indicate mVDC
17
.
4
Put the red test lead in V jack
4
, black one to the COM jack
3
5
Connect the red test lead with the positive of voltage which is waiting for measurement,
black one to the negative(ground)
6
The value of result show
16
*The number In the
, refering to the Explanation on Front Panel (Page 5)
DC V output
1
Press the Power switch
5
, turn on the Calibrator.
2
Press the input/output conversion key
8
, when the state of no output indicator
15
.
Make it under the state of output.
3
Press the V mV conversion key
6
,make it indicate VDC or mVDC
17
, at the range of output
you need.