Connect the red test lead with the positive of power which is outside, black one to the negative
of current which is waiting test.
7
If you want to change the output value or state, then press the value adjust key
9
,
10
,
11
,
12
or the mA mA% conversion key
7
.
Maintenance
Cleaning:
Periodically wipe the case with a damp cloth and detergent; do not use abrasives or solvents.
Calibration:
Calibrate your calibrator once a year to ensure that it performs according to its specifications.
Replacing the Battery
Please change the battery when the LCD indicates
.
Turn if the power of the Calibrator, when you change the battery, and screw off the breechblock on
the battery cabinet cover, then take off it and instead the fresh battery.
Replacing a Fuse
Warning!
To avoid personal injury or damage to the calibrator, use only a 0.125A 250V Fast fuse.
Fuse 1 is probably blown if:
In the V output mode, with the test leads removed from the calibrator, the display flashes OL.
Fuse 2: is probably blown if:
In the mA mode, the calibrator always reads 0.000, even with a signal applied.