background image

2. Wait for the display to show “0.0g” or “0.00g” then press and 

hold the [ MODE ] key for 3 seconds until the display shows 

“CAL” then release.

3. Press the [ MODE ] key again, and the LCD will flash the first 

required calibration weight. Place the required weight on the 

middle of the platform.

4. The LCD will flash then flash the second required calibration 

weight. Place the required weight on the middle of the 

platform.

5. The LCD will display “PASS” and return to normal weighing 

mode. Calibration is complete. You may now remove the 

weight and turn off the scale.

IV. TROUBLESHOOTING

The primary reasons for inaccuracy or malfunction are low 

batteries, incorrect calibration, overload or operating on an 

unstable surface. Please keep this in mind and maintain and 

operate your scale properly. This scale is a precise instrument 

and must be handled with the utmost care and caution.

[Display] Elimination of errors

[Lo] Change the batteries.

[OUT2] Recalibrate the scale as shown above.

[O_Ld] You are overloading the scale. Remove the excessive 

load immediately!

[UnSE] Use the scale in a more stable position