Avtron Model T477W Bonding Meter
with B14703 Battery Pack Replacement Instructions
Part Number T477WC15292
Operating Instructions
Subject to the EAR, ECCN [9E610.a].
This information is subject to the export control laws of the United States, specifically including the Export Administration Regulations (EAR),
15 C.F.R. Part 730 et. seq. Transfer, retransfer, or disclosure of this data by any means to a non-U.S. person (individual or company), whether in the United States or abroad, without any
required export license or other approval from the U.S. Government is prohibited, including without limitation any diversion to a military end user or use in a military end use application.
2-3
2.6
C15479 Probe Set (Optional)
These probes (Figure 2-2) have separate current and voltage leads, and the voltage sensing leads are bound together for nearly
the full length of the cable, which best minimizes the coupling effect described above. The best-rated meter accuracy is
obtained with this probe set when connected such that the meter reading is
positive
only. This probe set may be used for
calibrating the T477W Bonding Meter and whenever measurement accuracy is crucial.
NOTE
The red clamp and pencil style probe are positive current and voltage respectively while the black
clamp and pencil style probe are negative current and voltage respectively. The polarity signs on
the clamps have no meaning. Older probe sets current clamps and voltage pencil probes polarities
are determined solely by the lead color wired to them.
Figure 2-2.
Optional Probe Sets
C15292B1SCAN