Appendix - Battery Testing
Subject to the EAR, ECCN [9E610.a].
This information is subject to the export control laws of the United States, specifically including the Export Administration Regulations (EAR),
15 C.F.R. Part 730 et. seq. Transfer, retransfer, or disclosure of this data by any means to a non-U.S. person (individual or company), whether in the United States or abroad, without any
required export license or other approval from the U.S. Government is prohibited, including without limitation any diversion to a military end user or use in a military end use application.
7-2
7.
Batteries failing the discharge test may have a memory problem. A battery pack that has a memory problem may be
restored to a useable state by repeating the charge/discharge cycle described in steps 1 through 6. Usually batteries
require more than one recycle to regain normal capacity. Older batteries or others that have been mistreated may never
regain normal capacity but only recover to an acceptable capacity (≥50%). Those batteries which do
not regain an
acceptable capacity should be replaced.
8.
If battery pack is operating properly, reassemble the Bonding Meter.