TROUBLESHOOTING
If you obtain “strange” output waveforms, or unexpected values of t
RR
, keep these
points in mind:
1) The test jig output
must
be terminated with 50 Ohms.
2) The two coaxial cables connecting the test jig to the output module should
normally be 2 meters in length.
3) Keep device lead lengths as short as possible, to minimize parasitic inductance.
4) The test jig lid must be closed, or the pulser output will be disabled.
For technical support, contact
. Sample waveforms and digital
photos of your setup are always helpful!