FRONT PANEL CONTROLS
1. POWER Switch
. This is the main power switch. When turning the instrument on,
there is normally a delay of 10 seconds before anything is shown on the main
display, as the internal operating system boots up.
2. OVERLOAD Indicator
. When the instrument is powered, this indicator is normally
green, indicating normal operation. If this indicator is yellow, an internal automatic
overload protection circuit has been tripped. If the unit is overloaded (by operating at
an exceedingly high duty cycle or by operating into a very low impedance), the
protective circuit will disable the output of the instrument and turn the indicator light
yellow. The light will stay yellow (i.e. output disabled) for about 5 seconds after
which the instrument will attempt to re-enable the output (i.e. light green) for
abOUT1 second. If the overload condition persists, the output will be disabled again
(i.e. light yellow) for another 5 seconds. If the overload condition has been removed,
the instrument will resume normal operation.
This overload indicator may flash yellow briefly at start-up. This is not a cause for
concern.
3. SYNC OUT
. This connector supplies a SYNC output that can be used to trigger
other equipment, particularly oscilloscopes. This signal leads (or lags) the main
output by a duration set by the "DELAY" controls and has an approximate amplitude
of +3 Volts to R
L
> 50Ω with a pulse width of approximately 100 ns.
4. LIQUID CRYSTAL DISPLAY (LCD)
. This LCD is used in conjunction with the
keypad to change the instrument settings. Normally, the main menu is displayed,
which lists the key adjustable parameters and their current values. The
“Programming Manual for -B Instruments” describes the menus and submenus in
detail.
16
2
3
4
5
1
Содержание 14258
Страница 35: ...RST no query form SRE numeric value STB query only TST query only WAI no query form 35...
Страница 38: ...WIRING DIAGRAMS WIRING OF AC POWER...
Страница 39: ...PCB 158R6 LOW VOLTAGE POWER SUPPLY...
Страница 40: ...PCB 168C HIGH VOLTAGE DC POWER SUPPLY...
Страница 41: ...PCB 201C SAFETY INTERLOCK...
Страница 42: ...PCB 241F SECONDARY PULSE WIDTH OR DELAY CONTROL...
Страница 43: ...PCB 228D OP AMP LEVEL SHIFTING...
Страница 44: ...PCB 104H KEYPAD DISPLAY BOARD...
Страница 45: ...MAIN WIRING...
Страница 46: ...STANDARD TEST JIG WIRING AVX TRR NXPB SOD123W...
Страница 47: ...TEST JIG WIRING AVX TRR NXPB SOD128...
Страница 48: ...PERFORMANCE CHECK SHEET 48...