7800 Family User Instructions
17
5.2
The structure for these menus is shown in
. Specific information about each
mode followed by step-by-step instructions for accessing them are described in the
following pages.
Figure 5.3 Menu
Normal Weighing
Mode
Set Switch 1 to
the OPEN Position
Set Switch 2 to
the OPEN Position
Menu Mode
See note 1
See notes
See notes
See Note 5
Press
key
TEST
DIAG
CONF
9.xxxx
FILT
SLO
DISP
PROT
RA
BAUD
RO
CAP
I/0
TYPE
DIV-A
UNITS
LOC-GR
DIV-N
PRTOUT
CAL-GR
FILT
DONE
BAUD
Performs a
display test
Displays protocl
selection
Performs a
test of RAM
Performs a
test of ROM
Displays
capacity/
resolution
selection
Displays
baud/parity
selection
Performs a
input/output
test
Displays
rounding
type
Performs a
division test
w/AZT
Displays
unit of
measure
Displays
the local
gravity value
Performs a
division test
w/o AZT
Displays
output print
format
selection
Displays
the calibration
gravity value
Displays filter
selection
CAL
LBS
FAST
1000G
CLASS
Scale capacity is prompted.
See Table 1 for the list of
alternative weight available
for different capacities.
Returns to DIAG mode
If scale is placed in noisy
environment, set slow
filter to minimize the effects.
SCALE
LOAD 0
LOD
(CAP)
DONE
RE-CAL
ROUND
RESO
150.02
75.02
150.05
75.01
lb selections
kg selections
CLASS
Indicates Factory
Default Settings
Press the
key to move in the menu
ZERO
Press the
key to move in the menu
TEST
After making a selection, the display generally
moves to the next parameter in the list.
After choosing done, the display shows
DIAG.
Notes:
1) The 're-calibration' values displayed
will be the same as those of the
original 'calibration' units of measure.
2) Will flash
if originally
CAN'T
calibrated for a non-switching kg
capacity/resolution
3) Will flash
if originally
CAN'T
calibrated for a non-switching lb
capacity/resolution
4) When protocol is configured
for manual print (i.e. Print), all serial
input commands will be ignored.
5) To change the local gravity setting
Press and hold the
key to
ZERO
increase the value, or press and
hold the
key to decrease the
TEST
value. When done set Switch 2 back
to the CLOSED position.
50.01
100.02
60.02
150.02
75.01
150.05
75.02
250.05
100.02
300.1
150.05
PROT
8213
3835
SMA
AUTO-1
AUTO-2
NCI
PRTOUT
DONE
DONE
SCALE
Choose from a list
of 15 baud and
parity options
See Table 5.2
Choose from a list of 8
output formats. See Table 5.3