Min-Max Temperature Display:
The top left display (pictured to the right) is the minimum and
maximum temperature recorder. This displays the current air
temperature inside the cabinet, and also provides the user with the
minimum and maximum air temperatures reached inside the
refrigerator since the last reset.
To display the minimum air temperature
the cabinet has reached, since the last reset, press the
min
button on the Min-Max display, identified with a ‘down’ arrow. The temperature will flash for several
seconds, then the display will return to the current air temperature inside the cabinet.
Similarly,
to display the maximum air temperature
the cabinet has reached, since the last reset, press
the
max
button on the Min-Max display, identified with an ‘up’ arrow. The temperature will flash for
several seconds, then the display will return to the current air temperature inside the cabinet.
Please Note: The maximum temperature will rise if the door is left open for an extended period of time.
To reset the Min-Max display
, press and hold the red
set
button for approximately 5 seconds. The
display will show the number of hours that have been logged since the last reset, then
“rE”
will be
shown on the display to confirm all values have been reset.
It is essential that the reset button is pressed before leaving the premises to ensure a true
reading is viewed the following morning.
(Some regulations governing vaccine storage may require you to reset the display more frequently than
once daily, please check with your State/Territory health authority for further information).
The current temperature before pressing the re-set button should be between the required
temperatures of +2°C to +8°C.
The Min-Max Temperature Display is to be used as a guide only, providing an early indication of
developing conditions inside the refrigerator.
Always refer to your LogTag for an accurate record
of vaccine core temperature.
Reset Min/Max values