Avem Quirks
National Service Support Line:
1800 658 120
1
0
viii.
Set the logging interval to
Record a reading
every
3-5 minutes
. (The interval may be
adjusted to suit regulatory/insurance requirements).
ix.
Set the logger to
Begin recording after a delay
of
0 minutes
, to ensure the logger
begins recording temperatures immediately.
x.
Enable the
OK (Green)
and
Alert (Red)
indicators, ticking both boxes
i.
The Alert (Red) indicator should be enabled when readings are
below 2
or
above 8
°C. This will provide an additional alert to the operator, in the event
of a cold chain breach.
ii.
Tick the box to enable the Alert (Red) indicator after
2
consecutive readings
iii.
Tick the box to
Leave alert indicator enabled even if readings return
within alert range
.
iv.
Tick the box to
Clear and reset alert when START MARK button pressed
.
xi.
Leave the password fields blank, unless additional security is required for your recorder
configuration.
xii.
At the bottom of the window, click the NEXT button to progress to the next stage,
which will confirm the LogTag recorder is ready to be started. A confirmation message
(shown below in green) will be displayed on screen.
LogTag status confirmation