SAVe II+ OPERATOR’S MANUAL > Page 47
1
Ventilators set to simulate 5’9” patient with moderate ARDs
2
Operating the ventilator while it is charging will increase time to full charge
3
If stored at room temperature (15
– 20C) prior to use the ventilator will continue normal operation for up
to 1.5 hours at the lower temperature extreme of -20 C.
4
To maxim
ize the useful life of the device’s lithium-ion battery, store at temperature in lower range of the
specified long-term storage range with 40
– 60% of charge (2 battery lights)
5
Gas volumes are specified at ambient temperatures and pressure (ATP). Critical care ventilators often
quote the volume of the gas as it is in the patient’s lungs (BTPS) where the higher temperature and more
humid environment make the gas expand.
6
Allow up to 40 breaths to reach set PEEP.
7
Declared tolerance with 100% oxygen input. FiO2 should be adjusted if the medical gas supply or
oxygen concentrator outputs less than 100% oxygen. O2 concentrators and hospital supply lines
commonly output 93%.
8
Value depends on if the extendable airway connector is collapsed or fully extended. If ventilating with
low tidal volumes or adding accessories such as an HMEF it is recommended to collapse the extension
as much as possible to limit dead space in the circuit.