SAVe II+ OPERATOR’S MANUAL > Page 40
DEVICE
ALARM
The SAVe II+ software monitors multiple components to ensure they are operating
within expected parameters. The DEVICE alarm is triggered when device is outside of
specified temperature range OR a non-field correctible malfunction is detected.
Device responds by:
1) Activating “DEVICE” visual alarm indicator and high priority audible alarm
indicator.
2) Stops ventilating (cycling).
3) Opens the exhaust valve to allow patient to spontaneously breathe with minimal
resistance.
4) Displays a device error code in the TIDAL VOLUME (TV) display.
What to do:
If an error code is observed turn the device off, then on to clear transient alarms. If this
does not address the problem begin ventilating using another method.
See if the error code corresponds to one of the following potentially field correctable
issues:
-
E13
: Most likely caused by operating at temperatures below -10C. Increase device
temperature to above 0C.
-
E15
: Most likely caused by a loose battery connection or battery that has entered
a safe mode due to a malfunction. Verify the battery connection. If the device still
produces error code, then replace battery.
-
E16
: Most likely caused by operating at temperatures above 60C. Decrease device
temperature to below 50C.
For all other device codes or if the error is not field correctable then immediately take
the device out of service. Make a note of the error code and contact an authorized
service provider.