SAVe II+ OPERATOR’S MANUAL > Page 27
well so that it is not easily occluded. If the nipple becomes occluded, air will not be
delivered to the patient and a disconnect alarm will trigger. If the attenuator is exposed
to sand or dust wash it with water between uses. Make sure it is dry before reusing. In
a dust or sandstorm, the black air intake port cap will do a better job than an attenuator
at guarding against particulates getting into the pump.
Attach the attenuator by:
1)
Remove the black Intake Cap from air intake port. See Figure 12.
2)
Attach the attenuator to air intake port. Leave the debris filter in place
3)
To avoid blocking the nipple at the end of the attenuator, orient it so that is faces
into the port well
Figure 20: Attenuator Attachment
WARNING:
- Do not occlude small nipple as this will dramatically decrease tidal volume
- Leave the debris filter in place. You can see picture of debris filter in Figure 18
- In particularly sandy or dusty conditions use the air intake port cap instead of
the attenuator