SAVe II+ OPERATOR’S MANUAL > Page 44
Do not clean any portion of the SAVe II+ or its accessories with abrasives or chlorinated
hydrocarbon cleansers.
Do not allow dirt, sand, debris, grease, oil, or caustic chemicals to enter or coat the
unit or its accessories. To prevent debris from entering the SAVe II+, the Debris Filters
should always be securely in place and the port cover should be closed when the unit
is not in use. If the debris filter becomes saturated with dust or sand, turn the unit
upside down when removing the filter so any loose debris falls out rather than in the
unit. If sand or dust gets into the unit, service the unit before using as particulates may
significantly impact tidal volume.
Clean the port well and port cover prior to removing the debris filter. It is recommended
the SAVe II+ be stored in its carrying case when not in use.
WARNING:
- Sand or dust inside the pump may significantly decrease the volume of gas
delivered to the patient.
- Do not attempt to clean single-use accessories.
- To avoid damaging SAVe II+ plastic components and user interface, do not use
cleaning agents containing ammonium chloride, other chloride compounds,
more than 2% glutaraldehyde, phenols, or abrasive cleaners.
- Under no circumstances should the SAVe II+ or its accessories be immersed in
liquid. If the SAVe II+ becomes wet, the unit should be dried using a lint-free
cloth immediately, or once the unit is no longer in use. If the SAVe II+ becomes
immersed, discontinue use. Air dry with port door open and patient ports facing
down so water that may have entered the pump can drip out. Return to
appropriate service facility for inspection. DO NOT expose the switch, external
power jack, or audible alarm port directly to liquids.
- Never expose to an autoclave.
R
EPLACE
C
ONSUMABLES
Single-use accessories must be replaced after each use. Other accessories should be
replaced Replace the following after each patient use:
- Patient breathing circuit
- Debris filter
- Mask
- Heat and moisture exchanger (HME or HMEF)
Reordering information is available at http://www.automedx.com
BREATHING CIRCUIT
The SAVe II+ breathing circuit is single use. Examine the breathing circuit tubes for
cracking, discoloration, sharp edges, or other signs of damage. DO NOT attempt to
use or repair damaged breathing circuits. Damaged breathing circuits must be
replaced. If necessary, exterior walls of tubing may be cleaned with a damp cloth and