SAVe II+ OPERATOR’S MANUAL > Page 26
NOTE:
1) FIO
2
values assume the O
2
tank or concentrator is delivering 100% oxygen
2) Setting the oxygen source
flow rate higher than the minute ventilation of the
ventilator will unnecessarily deplete oxygen supply faster.
3) Specified FIO
2
values require a fully expanded oxygen reservoir tube. Failing to
fully extend the tube may materially decrease FIO
2
values for a given O
2
flow rate.
4) Many variables impact delivered FIO
2
values. If the concentration of delivered
oxygen is critical then it should be measured with a calibrated oxygen analyzer that
features a minimum and maximum concentration alarm.
WARNING:
- If using supplemental oxygen, avoid smoking or open flames. Leaks at oxygen
connections can cause dangerous O
2
levels in the vicinity of the leak. To avoid the
risk of ignition, visually inspect oxygen connections before and after connecting
supplemental O
2
and take measures to properly ventilate the area.
- Do not use oil, grease, or combustible lubricants (only those approved for oxygen use)
in contact with any part of the ventilator, regulator, or cylinder.
- Do not block the air intake port of the O
2
reservoir tube.
- Place end of oxygen reservoir tube in a location that will prevent sand or dust from
entering. The oxygen supply must be shut off when ventilation is interrupted.
- The line connecting the ventilator to the oxygen source must be designed exclusively
for use with medical-grade oxygen. Under no circumstances should the user modify
the low pressure O
2
line. In addition, the line must be attached without the use of
lubricants.
- Take required precautions when using oxygen. Do not use in explosive atmospheres
or near open flame.
N
OISE
A
TTENUATOR
If you are not using supplemental oxygen, consider attaching the noise attenuator to
the device to mitigate the sound. The reusable ATTENUATOR (P/N: M41112) is a U-
shaped tube designed to mitigate the noise of the device. The attenuator is connected
to the air intake port as shown in Figure 20. The attenuator is meant to be U-shaped.
Do not cut the fastener that holds the shape.
To keep the SAVe II+ small the attenuator has been developed as an external reusable
accessory rather than an internal component. It is not required for operation but can
be used to significantly dampen the noise. Remove the black air intake cap and
connect the female end of the attenuator over the air intake port, which contains the
debris filter (do not remove the debris filter) and position the nipple to sit inside the port